13. If the calibration was not successful, one of the error messages below will appear. Take the
indicated action to correct the problem, and then perform a new calibration.
"Err0" - The calibration test weight or the keyed-in weight is larger than the full capacity
of the scale. Change the calibration test weight or check the input data.
"Err1" - The calibration test weight or the keyed-in weight is smaller than 1% of the full
capacity of the scale. Change the calibration test weight or check the input data.
"Err2" – There is not enough signal from the load cells to establish a proper calibration.
Most commons causes include incorrect load cell wiring, a mechanical obstruction or a
faulty (damaged) load cell.
Error Messages
Indicates that the weighing capacity of the scale
has been exceeded. The weight limit of the
WB5040-SWP is 5000 pounds.
Indicates that the weight on the scale is negative.
Press the Zero/Off key.
Beam #1 not found; ensure that beam #1 is po-
wered ON and/or move closer to display unit.
Beam #2 not found; ensure that beam #2 is po-
wered ON and/or move closer to display unit.
Beam #1 batteries have been depleted and need to
be replaced.
Beam #2 batteries have been depleted and need to
be replaced.
Both Beam #1 and Beam #2 batteries have been
depleted and need to be replaced.
Display unit battery needs to be recharged