user manual
www.vetek.com
6
CAlIBRATIoN
We recommend the scale to be calibrated at yearly intervals.
The minimum test weight that can be used is 70% of full-scale capacity.
To AdjUST The ZeRo PoINT USING The f16 ZeRo AdjUST PRoCedURe:
.
1.
Turn the power off.
2.
Put Calibration Switch to the Calibration position.
3.
Turn the indicator oN.
4.
Scroll to “f 16”. Use TARe or oN to move up or down.
5.
Press ZeRo. The display will momentarily show “C 0” followed by a value. (If you
want, press dot to go back to parameter number.)
6.
After making sure that there are no weights on the platform, press ZeRo to zero out
the displayed value.
7.
Press NeT to save the zero point value. The display will show “endC0”momentarily,
and then revert back up to f16. At this time, proceed to the f17 span calibration to
complete indicator calibration.
To AdjUST The mAx PoINT USING The f17 SPAN AdjUSTS PRoCedURe:
1. Scroll to “f 17”, then press ZeRo to enter span calibration menu.
2. The display will momentarily show “C 1” for the span calibration, followed by a value with one flashing digit.
This value will be zero. Place the test weight on the Scale.
3. Pressing TARE or ON will change the position of the flashing digit.
4. Increase the flashing digit by pressing Dot. Decrease the flashing digit by pressing ZERO.
5. After setting the exact value, press NeT to save the value.
6. If the adjustment was successful, the display will show “endC1” momentarily, and then revert back up to f17.
“err0” – The calibration test weight or the adjusted keyed-in weight is larger than the full capacity of the
scale. Change the calibration test weight or check the input data.
“err1” – Change the calibration test weight or check the input data.
“err2” – The internal resolution of the scale is not high enough to accept the calibration value.
Select a larger parameter for the Span Gain (f2).
7. Put the Calibration Switch to the normal mode.
8. Check the Scale
ON
NET
GROSS
OFF
ZERO
TARE