![Vetek JCL-12 Service Manual Download Page 13](http://html.mh-extra.com/html/vetek/jcl-12/jcl-12_service-manual_1010065013.webp)
Vetek Weighing AB
www.vetek.com
JWL Service Manual
May not be reproduced in any form without written permission
PAGE
12
☆
The transmission rate is 9600
P04
Turn-on or off of
the weight check
memory
Automatic turn-on of the check
mode when the unit is turned on
next time.
☆
No automatic turn-on of the
check mode when the unit is
turned on next time.
P05
Parity check
☆
No parity.
The parity can be odd or even.
P06
Backlight selection
Backlight display will be on for
the whole process
No backlight display
☆
Backlight display will be on
when the weight is greater than
9 divisions.
P07
Printing mode
selection
Manual printing
☆
Stable printing (the next printing
will be made only when the
indication is returned to zero.)
Continuous printing
P08
Filter grade
selection
☆
Worst vibration resistance
Average vibration resistance
Good vibration resistance
Best vibration resistance
P09
Zero indication
range
☆
No restriction on the zero
indication