Vetek FR-MS User Manual Download Page 4

user manual

www.vetek.com

[email protected]

4

1. 

AOFF    

Time for auto off (120s / 180s / 240s /360s / off)

2.  

Unit

   Weight units: kg / lb

3. 

CAP

  capacity (1 ~ 9999 kg)

4.  

divS 

  Division (0,001 ~ 5 kg)

5.  

CAL

 

calibration weight (1 ~ 9999 kg)

6.  

trAn

   rS232 Transfer

7.  

t.b.r.

  rS232 Transfer band rate

8.  

Grav. 

  Gravity

9.  

trd. 

 

Transfer mode

10. 

BMI 

  WeIGHT/H2 ( on/oFF)

TABEL 2 – CALIBRATION WEIGHT

1. press and hold 

[ZERO] 

to turn the power on, and then press continuously 

[UNIT/WEIGHT] 

three times, 

the display will show the “calukg”.(press 

[UNIT/WEIGHT] 

key to selected the calibration unit)

2. press

 [ZERO] 

key, when the display will show aD value, press 

[ZERO]

 key, the display will show “SaVe” 

and then show calibration weight.

3. press 

[ZERO]

 key again, the display will show “SaVe” and then return to aD value. now calibration 

weight is complete.

BMI FuncTIon SeTTInG

1. Body on the pan, it should your weght: 80kg

2. When the display stable, press HOLD to keep the display 80kg

3. Press and hold the HOLD key, until it display HLT 100, it means to input your height in cm

4. Press HOLD and UNIT keys until the display 180, it means 180cm

5. Press TARE, it display 80kg and 25%.

THe MeanInG oF FuncTIon SeTTInG

Summary of Contents for FR-MS

Page 1: ...3 200 kg 0 05 kg 0 1 kg 0 2 kg 5 Power Rechargeable battery 6 AA AC adapter 9V 500mA 6 Power saving auto shut off timing selectable 120seconds 180seconds 240seconds 360secons Off 7 Operating Temperature 5 35 C 8 Operating Humidity 25 95 RH 9 4Keys HOLD BMI UNIT WEIGHT ZERO TARE HOLD BMI Press this key to store weight value UNIT WEIGHT Toggle among KG and LB unit ZERO Zero the weight TARE Tare the ...

Page 2: ... Three modes Normal mode Setting mode Internal Auto Calibration mode 19 RS232 function BESKRIVNING AV TANGENTERN HOLD BMI UNIT WEIGHT ZERO TARE 1 HOLD BMI A Press this key to store weight value B To release Hold feature Press the HOLD key again C To release Hold feature and tare the weight Press the TARE key D Press and hold this key the scale will enter to BMI setting 2 UNIT WEIGHT A Välj viktenh...

Page 3: ...user manual www vetek com info vetek com 3 Function table Press and hold HOLD BMI to turn the power and then press continuously UNIT WEIGHT three times the display will show the SET ...

Page 4: ...how the CALukg press UNIT WEIGHT key to selected the calibration unit 2 Press ZERO key when the display will show AD value press ZERO key the display will show SAVE and then show calibration weight 3 Press ZERO key again the display will show SAVE and then return to AD value Now calibration weight is complete BMI function setting 1 Body on the pan it should your weght 80kg 2 When the display stabl...

Page 5: ...isplay show GrAv press TARE the display show on press the HOLD the display show OFF When the display show ON means acceleration to gravity is actived When the display show OFF means acceleration to gravity is inactived 2 When acceleration to gravity is actived press ON OFF turn the power off then press ON OFF turn the power on then press UNITS key 3 seconds enter to acceleration to gravity mode 1 ...