manual
5
how to calibrate
**you must have an accurate the full load kilogram weight or combination of weights in order to
calibrate**
1. The scale must be powered OFF. Press and hold [ZERO] first, then Press [ON/OFF] while keeping
the [ZERO] button depressed, then release both keys. The display will show the a/D value (a series
of random numbers).
2. Wait for the stable indicator to be displayed, Press [UNIT] the display will show “0SaVE”, then the
display will return to the random a/D value.
3. Place the full load kilograms weight(s) on platform. Wait for the stable indicator to display, then
press [UNIT], the display will show “0SaVE”. Calibration is complete, remove the weight(s).
4. Turn the scale OFF (Press the ON/OFF key) , then turn it back ON and check some weight readings.
If calibration is still incorrect, repeat calibration but try it more slowly.
If calibration is repeatedly incorrect then: 1) Replace the batteries, 2) Try a more stable surface or
vibration and interference free location, 3) Be sure there is no static electricity on the scale (use an
anti- static spray to remove static electricity)
FEaTURES
␣ auto shut off (optional)
␣ auto calibration
␣ auto zero tracking
␣ aC Power (optional)
␣ Removable display w/3ft cord
␣ kg /lb conversion
␣ large lCD (6digits 15 mm high)
␣ hold function
␣ Large square weighing tray.
SPECIFICaTIONS
**SPECIal WaRNINg**
Cell-Phones, Cordless-Phones, and any radio-frequency device can cause temporary interference
and cause the scale to temporarily not work properly. Please do not use any electronic device near
the shipping scale. Just like in an airplane, do not use your cell-phone near the scale when it is in
use. While there is NO risk of permanent damage to the scale, interference can cause an incorrect
calibration or incorrect weight readings.
model number
Capacity
Division
FCS-30
30kg
10g
FCS-60
60kg
20g
FCS-100
100kg
20g/50g
FCS-150
150kg
50g/100g
Platter / Tray
330×320mm
Net/gross weight
4.1kg/4.5kg
Package
Standard carton: 44.5×34.6×7(cm3)
5 Units in one box: 46×36×39(cm3)
Operating
Temperature
Optimum: 15-30°C
Power source
6×aa Batteries or aC/DC
adapter 9~12V/100ma (optional)
www.vetek.
com
info@vetek.
com