HD453 GB Issue 08
25
14
Safety Inspection, Maintenance and Calibration
CALIBRATION MENU
1.
CAL TEMP
OK
→
CAL RH%
OK
→
RH CYCLE
OK
→
AMBIENT
OK
→
SAVE
OK
→
CANCEL
OK
→
T 37.3C
OK CAL!
RH 28%
OK CAL!
RHC 16s
OK - +
T 20.0C
OK - +
CALIBRATE THERMOMETER
.
TAKE READINGS AT SEVERAL
PLACES 40mm ABOVE THE BASE
OF THE EMPTY UNIT AND
CALCULATE THE AVERAGE.
CALIBRATE HYGROMETER.
PLACE HYGROMETER IN BASE OF
UNIT WITH NO WATER IN THE
RESERVOIR PAN.
HUMIDITY PUMP CYCLE TIME.
ONLY APPLICABLE WHEN USED
WITH THE BRINSEA ADVANCE
HUMIDITY PUMP.
AMBIENT TEMPERATURE
COMPENSATION.
AMBIENT TEMPERATURE USED IN
HEATER CONTROL ALGORITHM.
ONLY ADJUST IF TYPICAL ROOM
TEMPERATURE IS GREATER THAN
4°C (9°F) DIFFERENT.
BGHEAT
OK
→
BGH60%
OK - +
BACKGROUND HEATING LEVEL.
HEATER POWER ADJUSTMENT
USED IN HEATER CONTROL
ALGORITHM. ONLY ADJUST WITH
ADVICE FROM BRINSEA
PRODUCTS.
PRESS BOTH BUTTONS TO UNLOCK THE CALIBRATION MENU.
SELECT THE OPTION / RETURN TO THE MENU.
GO FORWARD ONE SCREEN / INCREASE THE VALUE.
GO BACK ONE SCREEN / DECREASE THE VALUE.