unit to the laser table.
Using the cable provided, connect the 8-pin circular Hirose connector to the D2-210 and

2.

connect the D-sub 9-pin connector to a free power connector on the D2-005 or other Vescent
electronics module. If you are using an alternate power supply, refer to the table above for
correct voltage input connections.
Connect the SMA connector to an oscilloscope.

3.

If you have the fiber-coupling option, connect a polarization maintaining fiber with FC/APC

4.

connector to the input fiber connector. At least 1 mW of optical power should exit the fiber.
Skip to step 9.

Note: When coupling the laser into the fiber, strive to
properly align the laser polarization to the polarization
axis of the PM fiber. Otherwise you will observe
significant temperature and stress induced birefringence
leading to input power fluctuations.

Open the top cover.

5.

Block the reference photodiode with a strip of index card.

6.

Adjust the light level by rotating the λ/2 waveplate to obtain about -1 V of signal.

7.

Shift the D2-210 about the mounting screws to maximize the light level.

8.

Set the laser to scan over the alkali transitions. If you observe spectroscopy, try to move the

9.

D2-210 to optimize the spectrum.
Tighten the 4 mounting screws.

10.

In about 20 min, a green LED should light (visible through the port next to SMA connector)

11.

indicating the final temperature has been achieved.
If the spectroscopy does not look similar to the plots shipped with the device, proceed to the

12.

Alignment Procedure below.
Use the Balance Adjust control to set the lock point to ground.

13.

Aligning the Spectroscopy Module

The spectroscopy module is shipped factory aligned and should not require alignment if ordered with
a matching D2-100-DBR laser or if ordered with the fiber input (-FC) option. If the unit was not aligned
to a laser at the factory, or if it falls out of alignment, the following steps can be used:

Fashion several 1 cm x 3 cm cards from an index card for use as viewing screens.

1.

Remove the cover on the Spectroscopy Module.

2.

If working with a Vescent D2-100 laser, follow the instructions in the 

DBR Laser Module

 section

3.

to adjust laser steering so the beam passes through the center of the input and output
apertures on the laser module.
Adjust the λ/2 waveplate on the input face to maximize the light power reflected by the input

4.

PBS.
If required, loosen the two 2-56 screws holding the tilt-plate assembly into the spectroscopy

5.

module and slide the assembly so that the input beam is centered on the PBS.
Loosen the two cone screws using a 0.050“ Allen wrench. Then tighten the cone screw farthest

6.

from the alignment rod until the brass ring just begins to grab. Tighten the second cone screw
similarly. The goal is when moving the PBS with the alignment rod, it moves easily, but with
some resistance.

Summary of Contents for D2-210

Page 1: ...greater than 5 MHz for photocurrents of 50 μA and above The high bandwidth of the feedback enables tight solid locking that is immune to vibrations and shock The D2 210 is powered by a cable that plugs into the power connectors on any D2 series electronics module The D2 210 is a significant upgrade to the D2 110 incorporating many suggestions from our customers The module is magnetically shielded t...

Page 2: ...h signal 5 MHz Noise 5 MHz 80 nV Hz Set Temperature 25 to 65 C Temperature Stability 0 1 C Beam Height 0 95 inches Total package Size L x W x H 3 6 x 5 x 1 7 inches Tab 2 Powering the D2 210 The D2 210 requires 5 and 15 VDC and ground to operate The power input is via a female 6 pin Hirose connector HR10A 7TR 6SA 73 Depending on your order configuration you should have received a power cable with t...

Page 3: ...2019 08 21 02 55 3 12 Spectroscopy Module Product Manuals https www vescent com manuals Fig 1 Beam path and component map ...

Page 4: ...pler subtraction option or passed directly to the reference photodiode The remaining portion is passed through a fixed λ 2 waveplate and PBS to split off the pump beam from the probe beam The pump beam is directed to the Output Optics assembly where it is combined with the probe beam on a PBS in a counter propagating configuration The pump beam is dumped through the reflection port of the PBS that sep...

Page 5: ...bers are marked on the connectors are listed below figure 3 below may be useful Fig 3 Pin numbering on female Hirose 6 pin connector Hirose Pin DB 9 Pin if applicable Input Voltage Max Current Draw Notes 1 4 and 5 Signal Power Ground Return path for power rail 2 8 7 V to 15 V 100mA See note below 3 9 7 V to 15 V 100 mA See note below 4 6 and 7 5 V 1 5 A 2 5 A 2 5 A max with Doppler subtraction opti...

Page 6: ...ium 4 5 40 C potassium 8 9 60 C Tab 4 Temperature Loop Gain Adj In newer units there is a middle trimpot for adjusting the gain on the PID temperature loops Higher gain clockwise will result in faster temperature settling until oscillation occurs Temperature Bias Adj Note this adjustment is only available on older D2 210 models For the Doppler subtraction option this control enables the user to in...

Page 7: ...eplate attached to the front plate by magnets can be used to select the splitting fraction of the Tilt Plate PBS Fig 5 Half wave plate adjuster Quick Setup Guide Set the D2 210 in the beam path of the laser so that the beam goes through the center of the 1 input and output holes Loosely apply the four 1 4 20 mounting screws and washers to affix the ...

Page 8: ...he 4 mounting screws 10 In about 20 min a green LED should light visible through the port next to SMA connector 11 indicating the final temperature has been achieved If the spectroscopy does not look similar to the plots shipped with the device proceed to the 12 Alignment Procedure below Use the Balance Adjust control to set the lock point to ground 13 Aligning the Spectroscopy Module The spectrosc...

Page 9: ...t use a small twisting motion To adjust the beam up and down use your other hand to turn the knurled cylinder Place a viewing card in the path of the pump beam as shown Center the beam through the 2 passage Move the viewing card to the output port of the pump probe PBS as shown below Adjust the 1 beam so that it is centered on the output face of the PBS At this point the beams should be well align...

Page 10: ...e sure the saturation beam dump is well centered on the output PBS and then check the quality of the spectroscopy The 2 56 screws holding the input and output optics assemblies can also be loosened to shift the optics assemblies around to improve the spectroscopy Tighten the cone screw farthest from the alignment rod Then tighten the last cone screw The 3 order helps If the spectroscopy changes th...

Page 11: ...2019 08 21 02 55 11 12 Spectroscopy Module Product Manuals https www vescent com manuals Fig 7 Non Doppler subtracted side and peak lock spectra Fig 8 Doppler subtracted side and peak lock spectra ...

Page 12: ...t com manuals doku php id d2 spectroscopy_module_210 https www vescent com manuals Printed on 2019 08 21 02 55 From https www vescent com manuals Product Manuals Permanent link https www vescent com manuals doku php id d2 spectroscopy_module_210 Last update 2019 06 25 22 53 ...

Reviews: