Purchase Includes
D2-125 Laser Servo
VPN00465 BNC cable (1ft)
VPN00457 Trimpot tweaker (large)
VPN00458 Trimpot tweaker (small)
VPN00456 DB9 Power Cable (1ft)
Options
The D2-125 can be purchased with two options:
-PL: Peak Lock option
-IP-X: Internal Power (where X=100 VAC, 120 VAC, or 230 VAC)
The Peak Lock option generates a 4 MHz dither signal that can be used to modulate the laser current.
The amplitude and phase of this modulation can be adjusted via front panel controls. The Error Signal
input will be demodulated relative to the dither signal which will generate a derivative of the input
Error Signal. This derivative signal will be used to lock the laser to a minima or maxima of the input
Error Signal. This is often referred to as peak-locking. If purchased with the Peak Lock option, the
Laser Servo can still be used in side-lock mode. A front panel switch controls whether the D2-125-PL is
in side-lock or peak-lock mode. When in side-lock mode the modulation / demodulation circuitry is
disabled.
In this manual, sections that are only relevant to the Peak Lock option are printed in red.
The Internal Power option removes the 9-pin power connect used to connect power between Vescent
electronics modules and provides power entry module for powering the module via main AC lines. The
input voltage, 100VAC, 120VAC or 230VAC, is specified in the option. Note, with this option, the
D2-125 cannot be powered via the D2-005 Power Supply, the D2-001 break-out or from any other
Vescent electronics modules. Additionally, this module will not be able to power other Vescent
electronics modules; the -IP option makes the Laser Servo completely stand-alone.
In this manual,
sections that are only relevant to the Internal Power option are printed in orange.
Absolute Maximum Ratings
Note: All modules designed to be operated in laboratory environment
Parameter
Rating
Environmental Temperature
>15°C and <30°C
Environmental Humidity
<60%
Environmental Dew Points
<15°C
When connecting Servo Output to your
plant, always make sure to discharge any
residual voltages, for instance from a PZT.