background image

 
 

7.  There are three different ways to handle the reporting of volume. 

 

Do not calibrate drop volume

 In this case, you can skip to Step 8, and begin 

collecting data.  

 

Manually enter a drops/mL ratio value

 If you have done a previous 

calibration with a particular apparatus, you can enter the drops/mL ratio.  

 

Perform an Automatic drops/mL calibration

 You can use a 10 mL graduated 

cylinder and perform this calibration (see “To Calibrate Drop Volume” section 
below for specific directions on this procedure).  

 

8.  Start data collection. 

Important:

 No data will be collected until the first drop 

falls through the Drop Counter’s slot. Completely open the lower 2-way valve 
(the upper valve should still be set to a slow drop rate). You will now see pH vs. 
volume data being plotted on a graph. When you are satisfied that the titration has 
proceeded past the equivalence point, stop the data collection, and shut off the 
lower 2-way valve of the reagent reservoir. A plot of pH 

vs

. volume will be 

displayed for your data collection.  

9.  After collecting data, the program will allow you to view graphs of the first or 

second derivative 

vs.

 volume—second derivative 

vs.

 volume is shown here. Data 

from calculators and Palm OS handhelds can be imported into Logger 

Pro

 3 for 

further analysis or enhanced display or printing capability. 

To Calibrate Drop Volume (in mL) 

To calibrate drops so that the volume of titrant is recorded in units of milliliters, 
choose Calibrate Drops in your program. The Automatic menu choice helps you to 
measure the volume of an individual drop by counting the number of drops that pass 
through the Vernier Drop Counter and dividing by the total volume of the drops. To 
calibrate drops: 

a.  Place a 10 mL graduated cylinder below the slot of the Drop Counter. 

b.  Fill the reagent reservoir with titrant. 

c.  Choose to start the calibration routine in your program. 

d.  Open the bottom 2-way (on-off) valve to begin releasing drops from the 

vessel through the Drop Counter (the top valve finely adjusts the drop rate). 

e.  Continue to release drops until there are between 9 and 10 mL of liquid in 

the graduated cylinder. 

f. 

Close the bottom valve on the vessel to stop the drops. 

g.  Type the precise volume of liquid in the graduated cylinder in the Volume 

(mL) box of the Calibrate drops dialog box. 

h.  Choose to stop the calibration routine of your program. 

i. 

The number of drops/mL will be displayed. It may be helpful to record the 
value for that particular reagent reservoir for future titrations. 

j. 

You can now continue with the titration (return to Step 8 above). 

Alternately, the Manual choice allows you to input a previously determined value for 
drops/mL. Simply enter the drops/mL value (e.g., 28.0), then continue with the 
titration. 

 
 

6

To Calibrate the pH Sensor (or a different sensor) 

For many titrations, it will not be necessary to calibrate the pH Sensor. Without 
calibration, you can expect pH readings to be accurate to ±0.25 pH units. Having pH 
offset by this much will not affect the location of the equivalence-point volume. In 
those cases where a very accurate calibration is required, however, our programs 
allow you to do a two-point calibration, using two pH buffer solutions. The general 
procedure, using a pH-4 buffer and a pH-7 buffer as an example, is: 

 

First Calibration Point 

a.  In your program, choose to do a new calibration. 

b.  Rinse the pH Sensor with distilled water and place it into a buffer of pH 4.0. 

c. Type 

4

 in the edit box as the pH value. Swirl the sensor, wait until the 

voltage stabilizes, then choose to keep the calibration point. 

Second Calibration Point 

d.  Rinse the pH Sensor with distilled water, and place it into a buffer of pH 

7.0. 

e. Type 

7

 in the edit box as the pH value for the second calibration point. 

Swirl the sensor and wait until the voltage stabilizes. Click Keep, then click 
OK. This completes the calibration. 

f. 

You are now ready to collect data using the new calibration. 

Note: 

If you 

want to use these new calibration values at a later time, you can simply 
record the slope and intercept values found on the Manual Entry tab and 
type them in when needed. 

The Manual Entry tab allows you to enter the Slope and Intercept values determined 
during a previous two-point calibration. 

Reviews: