2.  Assemble your equipment as shown here.  

a.  Note that the magnetic stirrer, ring stand, 

100 mL beaker, and utility clamp are not 
included with the Vernier Drop Counter. As an 
alternative to using the magnetic stirrer, 
students can use a stirring rod to stir manually 
(we do not recommend this as a standard 
method). 

b.  Arrange your magnetic stirrer on the base of 

the ring stand. Place a 100 mL beaker on the 
center of the stirrer. 

c.  Remove the pH storage bottle from the pH 

Sensor. Insert the pH Sensor body through the 
larger round hole on the Vernier Drop 
Counter.  

d.  Slip the Microstirrer (shown here) onto the bottom 

of the pH Sensor. Slide the Vernier Drop Counter 
down the ring stand to a level such that the 
Microstirrer is very close to the bottom of the 
beaker. Tighten the turn screw of the Drop Counter 
to hold it firmly in place. 

e.  Connect the pH Sensor to CH 1 of the interface, and connect the cable of 

the Vernier Drop Counter to the 

DIG/SONIC 1

 channel. 

f. 

Connect the spout and two 2-way valves to the plastic reagent reservoir. 

Note:

 There are 

two

 2-way valves below the reagent reservoir. In these 

directions, we will use the 

bottom valve

 as an on-off valve (either 

completely open or completely shut). We will use the 

top valve

 as an 

adjustment valve, to deliver drops at a slow, consistent rate. 

3.  Making sure that both 2-way valves are in the 

closed position (horizontal), add about 20 mL of 
titrant to the plastic reagent reservoir. For your 
first run, we recommend using 0.1 M NaOH 
solution.  

4.  Before collecting data, or calibrating the drops, 

you should adjust the flow rate of the two 
valves of the reagent reservoir. Temporarily, 
place another beaker below the spout of the 
reagent reservoir. First, completely open the 
bottom 2-way valve; then slowly open the top 
valve until a very slow drip rate is achieved–a 
rate of one drop every two seconds, or slower. 
Now close the bottom valve. 

 
 

4

5.  Add the solution to be titrated to the 100 mL beaker. For this run, we recommend 

using about 5 mL of 0.1 M HCl solution. Add enough distilled water so the 
solution level covers the bulb of the pH Sensor (about 35–40 mL). If you are 
going to use a small magnetic stirring bar, or use a stirring rod, less distilled 
water will be required (about 1–20 mL). Optional: Add two drops of 
phenolphthalein indicator. Turn on the magnetic stirrer. 

Important: Minimizing 

volumes of solution titrated, and maximizing stirring efficiency will 
significantly improve your titration results! 

 

6.  Start up the appropriate program.

1

 

Computer (Using Logger 

Pro

 3 software) 

If using Logger 

Pro

 3 software, open the file “Drop Counter-pH” from the Drop 

Counter folder. The Drop Counter folder is located in the Probes & Sensors folder. 

Computer (Using Drop Counter software) 

If using the Drop Counter software on a computer, simply launch the program.

2

 

LabQuest 

If using a LabQuest, start the LabQuest App. If your sensor does not auto-ID, 
manually set up the sensor. 

TI Graphing Calculator 

If using a calculator, start the DataDrop program or the EasyData App. 
Follow the steps below to use the DataDrop program. 
a. Press 

the 

 key on your calculator and select DataDrop.  

b.  Use the numbered options (1-6) to prepare to collect data. If you wish to use 

the stored calibration for the Drop Counter, type 

2

 to start collecting data. 

Follow the steps below to use the EasyData program. 
a. Press 

the 

 key on your calculator and select EasyData.  

b. Select 

8

 (Other Sensors…) from the Setup menu. 

c. Select 

5

 (DIG/SONIC1) as the location of the Drop Counter. Select 

d. Select 

2

 (Drop Counter) from the Select Sensor menu. Select 

 

Palm OS Handheld 

If using a Palm OS handheld, you will need to perform a few setup steps. 
a.  Start Data Pro from the Applications screen.  

b.  Choose New from the Data Pro menu. 

c.  On the Main screen, tap 

d.  Tap the DIG/SON 1: button and choose Drop Counter. 

e. Tap 

 to return to the Main screen. 

 

                                                 

1

 DataPro will stop collecting data at 501 points, which is approximately 20 mL of 

titrant added. 

2

 Logger 

Pro

 3 will stop collecting data a 1500 points, which is approximately 

60 mL of titrant added. 