35
Water
leaking
internally
Mechanical seal
leaking.
Check if water is leaking from the motor
drain hole - Replace damaged seal.
Check the water feed pipe connection
to the seal – Repair.
Leaking pipe work.
Lid arm leaking.
Motor flange leaking.
Check diaphragm sealing joint –
Replace.
Check pipe clips – Tighten.
Check condition of pipes – Replace.
Check joint lid arm inlet pipe – Replace
pipe if cracked / tighten clip.
Check lid arm sealing inlet block seal to
arm moulding – Re seal joint.
Check gasket on motor flange –
Replace.
Lid seal
leaking.
Bubbling
Lid seal is not sealing.
Debris on the seal face – Clean.
Check lid arm alignment – see fault 4.
Hinge bushes worn – Replace.
Check condition of lid seal – Replace.
Check operation of lid handle insert –
Repair.
Check seal is seated correctly - Replace
Leaking pipe work.
Pressure transducer not
operating.
Back pressure from
adjacent items in
Sluice room.
Hopper vent outlet blockage – Clean.
Vent pipe blockage – Clean.
Check the press switch operation –
Replace.
Check for water in the pressure switch
pipe – Clean & dry.
Check that there is no blockage in the
service drain – Clean.
Check that all equipment served by the
service drain is installed correctly –
Amend if necessary.
Fault 6
Fault 7
Summary of Contents for Vortex
Page 1: ...Issue No 2 ...
Page 8: ...8 Figure 1 Dimensions of the Vernacare Vortex ...
Page 20: ...20 7 Lid Arm Assembly Figure 5 Lid Arm Assembly ...
Page 25: ...25 11 Water Distribution Figure 9 Water Distribution ...
Page 27: ...27 12 Electrical Circuit Figure 10 Electrical Circuit ...
Page 42: ...42 The latest version of this manual can be found at www vernacare com ...