Calibration test Unit for Personnel 

GroUndinG testers

installation, operation and Maintenance

UNIT C, 4TH DIMENSION, FOURTH AVENUE, LETCHWORTH, HERTS, SG6 2TD UK

Phone: 0044 (0) 1462 672005, Fax: 0044 (0) 1462 670440 • e-mail: 

[email protected]

, Internet: 

Vermason.co.uk

tb-7578

 August 2010 Page 1 of 2 

© 2010 Vermason.

operation

The Vermason 

223001

 Calibration 

Test Unit is designed to simplify 

the process of calibrating wrist 

strap and footwear testers. The 

calibration unit allows the user to 

quickly and easily verify whether 

a tester is operating within 

specifications. This product can 

be used as one of the tools to 

fulfill the EN 61340-5-1 clause 

5.2.3 “Compliance Verification 

Plan”. “Compliance verification 

records shall be established and 

maintained to provide evidence 

of conformity to the technical 

requirements.  The test equipment 

selected shall be capable of 

making the measurements 

defined in the Compliance 

Verification Plan.”

The Vermason 

223001

 Calibration 

Test Unit is a passive device, 

and requires no power source. 

The calibration test unit is 

manufactured to check the test 

ranges for both wrist strap and 

person/footwear systems per EN 

61340-5-1 Table 2 Personnel 

Grounding of less than 3.5 x 10E7 

ohms upper limit and detecting 

the presence of current limiting 

resistance of  5 x 10E5 ohms. 

The 

223001

 is supplied calibrated 

to NIST traceable standards.

The Vermason 

223001

 Calibration 

Test Unit is an important tool for 

optimization of an ESD control 

program. A leading corporation 

estimates a $95 to 1 return on 

every dollar invested in ESD 

control. However, the ESD 

protective products utilized must 

be working, so it is essential that 

testers are in calibration when 

used to verify the parameters of 

wrist straps and ESD footwear. 

The Vermason 

223001

 Calibration 

Test Unit easily pays for itself. 

Instead of returning testers 

to Vermason, a Calibration 

House, or even your Metrology 

Department, testers can be 

calibrated where located in the 

factory being out-of-service only a 

couple minutes. There is no need 

to have an extra replacement 

Tester. We recommend a 

minimum of annual calibration.

Monitoring and control 

instruments are out of the 

scope of RoHS (Restriction of 

Hazardous Substances).

to use the calibration 

unit:

1. Fit the two cords into the 

sockets on the calibration unit.  Fit 

the other end of one lead into the 

socket on the top of the unit.

2. Turn the selector dial to the 

required test setting (usually fail 

high or low).

3. Press the end of the remaining 

lead onto the operating plate of 

the unit. 

teCHniCal bUlletin tb-7578

Made in Britain

Figure 1. 

223001

 Calibration Test 

Unit.

4. The relevant LED should light, 

and beeper should sound if it is a 

pass.

5. Repeat for the other settings on 

the calibration unit.

Calibration of the 223001

instruments required:

BM80 Mega-ohm tester

2 x 4mm Plug leads

instructions to test:

Connect the BM80 to the tester 

using the 4mm leads.  Switch the 

BM80 pointer to 50V test.

Point the rotary switch on the 

H330 Fail low to Fail high one at a 

time and press the button on the 

BM80 until meter gives a stable 

reading.

Record all the reading and 

compare with the following:

Fail Low: 

793.8 – 826.2 

 

 

kilohms

Pass Low: 

970.2 – 1009.8 

 

 

kilohms

Pass high: 

30.87 – 32.13 

 

 

megohms

Fail high: 

37.73 – 39.27   

 

 

megohms

 

If all the recorded readings are 

within the range above, the unit 

can be certified and labelled as 

tested. 

Vermason’s 

223001

 Calibrates 

items 

222504

 and 

222506

.