Table 201
:
Guidance and manufacturer’s declaration
- electromagnetic emission
The NeuroTrac® product is intended for use in the electromagnetic
environment specified below. The customer or the user of the The NeuroTrac®
product should ensure that it is used in such environment
Emission test
Compliance
Electromagnetic environment
guidance
RF emission
CISPR 11
Group 1
The NeuroTrac® product uses RF energy only
for its internal function.
Therefore, its RF
emissions are very low and are not likely to
cause any interference in nearby electronic
equipment
RF emission
CISPR 11
Class B
The NeuroTrac® product is suitable for use in
all establishments , including domestic
establishments and those directly connected
to the public low voltage power supply
network that supplies buildings used for
domestic purposes
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Not applicable
Table 202
:
Guidance and manufacturer’s declaration
– electromagnetic immunity
The NeuroTrac® product is intended for use in the electromagnetic environ-
ment specified below. The customer or the user of theNeuroTrac® product
should assure that it is used in such an environment, and that precautions
regarding that environment are heeded.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood, con-
crete or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30 %.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic
fields should be at levels char-
acteristic of a typical location in
a typical commercial or hospital
environment.