5.0 Operation
5.4
Atmospheres
When an optional gas inlet is fitted, there is a label near the inlet saying "INERT GAS
ONLY". In practice, inert or oxidising gases may be used, but not combustible or toxic
gases.
The chamber is not gas tight, the gas usage may be high and the chamber is always
likely to contain some air. Residual oxygen of approximately 1% to 2% is to be
expected.
5.5
Operator Safety
This product incorporates a safety switch which interrupts the heating
element circuit when the furnace is opened. This prevents the operator
touching a live heating element and also prevents the product from
heating up if the furnace is left open. The operation of this switch should
be checked periodically.
Depending on use, the surfaces in the working chamber and the
chamber load may still be very hot after the appliance is switched off.
Touching these surfaces may cause burns. Use suitable personal
protective equipment or wait until the appliance cools down to ambient
temperature.
Before removing a hot object from the product, make sure there is a safe place to put it
down. If necessary use tongs, face masks and heat resistant gloves. Heat resistant
clothing and face protection can guard against the effects of radiated heat when the
furnace is open.
When the product is opened during operation there is considerable radiated heat. Do
not keep any flammable objects near the product, nor objects which could be damaged
by radiated heat.
5.6
Solenoid Valve with Manual Switch (if fitted)
If ordered the solenoid valve is operated using the panel mounted
switch. When the switch is in the 'ON' position the solenoid valve will
allow gas to
flow. Ensure the installation and use of the product does not create a hazardous
atmosphere. The workspace must have sufficient ventilation.
14
Summary of Contents for Carbolite Gero WV-11-01
Page 28: ......
Page 29: ...Notes Service Record Engineer Name Date Record of Work ...