When connecting the product to the power supply, the plug or isolating switch should
be accessible, easy to remove, and within reach of the operator.
The supply MUST incorporate an earth (ground).
Supply
Terminal Label
Cable Colour
Supply Types
Live - Neutral
Reversible or Live-
Live
1-phase
L1
Brown
to live
to either power
conductor
(For USA 200-240V,
connect L1)
N / L2
Blue
to neutral
to the other power
conductor
(For USA 200-240V,
connect L2)
PE
Green/ Yellow
to earth
(ground)
to earth (ground)
Electrical Connection Details:
3-phase L1
Black
to phase 1
L2
Black
to phase 2
L3
Black
to phase 3
N
Light Blue
to neutral (
if fitted
)
PE
Green/ Yellow
to earth (
ground
)
DO NOT connect a product ordered for three phase use to a single
phase supply or to the wrong type of three phase supply.
Technical Notes
This model has an electronically controlled power limit. Do not attempt to calculate the
Amps from the Wattage. High break capacity fuses should be used. Avoid fast-blow
fuses and magnetic trip circuit breakers - consult Carbolite Gero if in doubt.
Two-phase models use two phases out of a 3-phase + neutral supply. One phase is
unused. The neutral current is equal to the current in one of the used phases.
9