5.2.1
Cleaning
The product's outer surface may be cleaned with a damp cloth. Do not allow water to
enter the interior of the case or chamber. Do not clean with organic solvents.
Under no circumstances should any objects be placed on top of the
product. Always ensure that any vents on the top of the product are clear
of any obstruction. Always ensure all cooling vents and cooling fans (if
fitted) are clear of any obstruction.
5.3
Calibration
After prolonged use, the controller and/or thermocouple may require recalibration. This
is important for processes that require accurate temperature readings or for those that
use the product close to its maximum temperature. A quick check using an
independent thermocouple and temperature indicator should be made from time to
time to determine whether full calibration is required. Carbolite Gero can supply these
items.
Depending on the controller fitted, the controller instructions may contain calibration
instructions.
5.4
After-Sales Service
Carbolite Gero Service has a team of Service Engineers who can offer repair, calibration
and preventive maintenance of furnace and oven products both at the Carbolite Gero
factory and at customers’ premises throughout the world. A telephone call or email
often enables a fault to be diagnosed and the necessary parts to be despatched.
In all correspondence please quote the serial number and model type given on the
rating label of the product. The serial number and model type are also given on the
back of this manual when supplied with the product.
Carbolite Gero Service and Carbolite Gero contact information can be found on the
back page of this manual.
5.5
Recommended Spare Parts and Spare Parts Kit
Carbolite Gero can supply individual spare parts or a kit of the items most likely to be
required. Ordering a kit in advance can save time in the event of a breakdown.
Each kit consists of a thermocouple, a solid state relay, an instrument switch, an
element or set of elements and a door seal; for fan models the kit includes a fan and
motor assembly. Individual spare parts are also available.
When ordering spare parts please quote the model details as requested above.
5.6
Power Adjustment
The control system incorporates electronic power limiting, but for the model listed in
this manual the power limit is set to 100%. The power limit parameter OP.Hi may be
accessible to the operator, but should not generally be altered.
17