Product without supply cable: a permanent connection to a fused and isolated supply
should be made to the internal terminals after temporary removal of the product
access panel. The access panel is located at the top of the 'A' models and the right
hand side of the 'B' models.
Connection by line plug: the plug should be within reach of the operator and should be
easy to remove.
Connection to isolating switch: this should operate on both conductors (single phase)
or on all live conductors (three phase), and should be within reach of the operator.
The electrical supply MUST incorporate an earth (ground).
Supply
Terminal Label
Cable Colour
Supply Types
Live - Neutral
Reversible or Live-
Live
1-phase
L1
Brown
to live
to either power
conductor
(For USA 200-240V,
connect L1)
N / L2
Blue
to neutral
to the other power
conductor
(For USA 200-240V,
connect L2)
PE
Green/ Yellow
to earth
(ground)
to earth (ground)
Electrical Connection Details:
3-phase L1
Black
to phase 1
L2
Black
to phase 2
L3
Black
to phase 3
N
Light Blue
to neutral (
except delta
)
PE
Green/ Yellow
to earth (
ground
)
DO NOT connect a product ordered for three phase use to a single
phase supply or to the wrong type of three phase supply.
2.4
Fan Rotation Indicator
Note:
Do not connect an oven ordered for three phase use to a single phase supply
or to the wrong type of three phase supply.