5.5
Power Adjustment
The control system incorporates electronic power limiting, but for the model listed in
this manual the power limit is set to 100%. The power limit parameter OP.Hi may be
accessible to the operator, but should not generally be altered.
In some cases the supply voltage may be outside the range 220-240 V or the 3-phase
equivalent, the power limit parameter may be set to a value other than 100%. Do not
increase the value to 100%, see section 9.0 for details of power limit settings.
15
Summary of Contents for CARBOLITE GERO LGP 7/730
Page 24: ......
Page 25: ...Notes Service Record Engineer Name Date Record of Work ...