9.0
Fuses and Power Settings
9.1
Fuses
F1 - F2: Refer to the circuit diagrams.
F1
Internal
Supply
Fuses
Fitted if supply cable fitted.
Fitted on board to some types
of EMC filter.
GEC Safeclip of the type shown
(glass type F up to 16 A)
38 mm x 10 mm type F fitted on
EMC filter circuit board(s)
F2
Auxiliary
Circuit Fuses
Fitted on board to some types
of EMC filter.
May be omitted up to 25 Amp/
phase supply rating.
2 Amps glass type F
On board: 20 mm x 5 mm
Other: 32 mm x 6 mm
Customer
Fuses
Required if no supply cable
fitted.
Recommended if cable fitted.
See rating label for current;
See table below for fuse rating.
Model
Phases
Volts
Supply Fuse Rating (Amps)
F1
EZS-3G 12/600
1-phase
200-208
16 A
EZS-3G 12/600
1-phase
220-240
16 A
9.2
Power Settings
The power limit settings (OP.Hi) for this model is as follows. The figures represent the
maximum percentage of time that controlled power is supplied to the elements. Do not
attempt to "improve performance" by setting a value higher than the one from the
table. To adjust the parameter refer to the "Changing the Maximum Output Power" of
the control section of the manual.
Model
Volts:
208 V
200 V
220 V
230 V
240 V
EZS-3G 12/600
(1-phase or 3-
phase)
100/100/100
100/100/100
89/89/89
81/81/81
75/75/75
33