8.0 Fault Analysis
B.
Product Overheats
1.
Product only heats up
when the instrument
switch is ON
The controller
shows a very high
temperature
The controller is faulty
The controller
shows a low
temperature
The thermocouple may be
faulty or may have been
removed out of the heating
chamber
The thermocouple may be
connected the wrong way
around
The controller may be faulty
2.
Product heats up
when the instrument
switch is OFF
The SSR has failed
"ON"
Check for an accidental wiring
fault that could have
overloaded the SSR
36
Summary of Contents for CARBOLITE GERO CFM 14/2 AUX
Page 4: ...User Power Setting Adjustments 40 11 0 Specifications 41 11 1 Environment 41 4 ...
Page 6: ...1 0 Symbols and Warnings Caution Double Pole Neutral Fusing 6 ...
Page 20: ...3 0 2416 Controller 3 6 Navigation Diagram 20 ...
Page 37: ...9 0 Wiring Diagrams 9 1 WB 12 10 AUX 9 2 WB 12 11 AUX 37 9 0 Wiring Diagrams ...
Page 38: ...9 0 Wiring Diagrams 38 ...
Page 42: ......
Page 43: ...Notes Service Record Engineer Name Date Record of Work ...