Venner DruGuardian Operating Manual Download Page 6

Close the drug cabinet door, ensuring no penetration of light. The Venner 

DruGuardian

 is now active. 

To test the device is working, open the drug cabinet door and ensure no 

movements in front of the cupboard. After a delay the alarm should trigger.  

To stop the alarm either move in front of the cupboard or close the door.

During normal operation Venner DruGuardian

 will monitor and record activity 

in front of the open drug cabinet. If the drug cabinet is abandoned for a period of 

time (determined by the alarm delay rotary switch), the alarm will initially emit 

an intermittent beeping sound, progressing to a constant beeping sound until 

movement is detected. 

Figure 9: Venner DruGuardian

 loaded in cradle for below or above shelf

Lock the Venner DruGuardian

 into the cradle, remove the key and store it 

in a safe place. All devices in the same institution will be accessible with the 

same key.

CAUTION: Only the nominated person for the institution should have access 

to the key.

Figure 8b: Venner DruGuardian

 and cradle if affixing above shelf

In addition to door opening and closing times the data micro SD card will record 

when the alarm progressed from intermittent to constant sound. Captured 

images from the camera will be stored on the image micro SD card along with a 

time stamp and location cradle RFID code reading. 

Contents from both micro SD cards can be uploaded onto  

www.VennerDruGuardian.com

 to generate a range of reports for analytical 

purposes. Visit the website for detailed instructions.

Replacing the micro SD cards: 

(i) Remove Venner DruGuardian

 from the cradle. Be sure to note down the 

cradle RFID code and the physical location of the cradle, in order to associate 

the data downloaded from each Venner DruGuardian

 later. This should be 

stored with the nominated person/key holder.
(ii) Remove one or both micro SD cards – each card will contain the 12-digit 

RFID code from the fixed cradle. If the same SD card is shared between multiple 

cradles, it will contain all RFID codes from the fixed cradles according to the 

sequential arrangement it had been in.
(iii) Replace the micro SD cards making sure the cards are unlocked if appropriate.

There are two ways of charging the Venner DruGuardian

:

1.

   Using the mains charger or USB charger from your Venner DruGuardian

 

supplier:

• 

Remove the Venner DruGuardian

 from its cradle

• 

Set all 8 function switches to OFF (downwards)

• 

Connect the mains charger to the socket at the rear of the unit or to  

the USB socket

• 

After charging for 16 hours, remove the charger, and it is now ready for use 

(refer to Section 2.3)

Battery and charging 

3.

Figure 10: Mains charger socket

 11

 10

Summary of Contents for DruGuardian

Page 1: ...Medical Singapore Pte Ltd 35 Joo Koon Circle Singapore 629110 www vennermedical com Venner Medical Technologies SA Osprey House Old Street St Helier Jersey JE2 3RG www vennermedical com en OPERATING...

Page 2: ...s 5 2 1 1 Alarm delay rotary switch 6 2 1 2 Function switches 6 2 1 3 Security level switch 8 2 2 LED battery indicator 8 2 3 Using the device 9 3 Battery and charging 11 4 Service and maintenance 12...

Page 3: ...exposure to light Venner DruGuardian records detailed time and alarm data while the wide angle camera captures images in front of the Venner DruGuardian facilitating identification of persons in the v...

Page 4: ...photos from the camera are stored on the image micro SD card 3 Data log Door opening times and alarm data are stored on the data micro SD card No information is stored on the data micro SD card 4 Vol...

Page 5: ...y Three security levels low medium or high may be set based on user requirement The battery indicator illuminates different LED colours which represents CAUTION Recharge the device when the Dim or Sol...

Page 6: ...camera will be stored on the image micro SD card along with a time stamp and location cradle RFID code reading Contents from both micro SD cards can be uploaded onto www VennerDruGuardian com to gene...

Page 7: ...nnocent persons VMSG will not be held responsible should incidents occur VMSG will not provide any guarantee of continuous uninterrupted function of the device or that images will be captured in an id...

Page 8: ...t may contain materials which if disposed with general waste could be damaging to the environment Residents of the European Union must follow specific disposal or recycling instructions for this produ...

Page 9: ...Venner Group of Companies This product is covered by patents and patent applications of the Venner Group of Companies in the US and other territories Visit www vennerdruguardian com intellectual prop...