background image

 

The rotation direction set has its relative LED lighted on: clockwise (C) and anticlockwise (E) 

Torque 

 

The torque indicator bar (B), made by four LEDs, shows the intensity of the torque applied 

 

CONTROLLER Digital model does

n’t show a precise torque value as the Advance model 

Lock 

 

Holding the Lock key (H) for 

3 seconds it’s possible to lock the control panel 

 

Unlock the control panel by holding key H for 3 seconds 

 

When the instrument is locked LED (I) lights up 

 

 

4.  External Connections  

USB 

 

This model has USB connection in the back for software upgrading by PC 

5.  Menu 

Hold the control knob (F) for 3 seconds to enter into the setting mode when stirring is off. 

Turning the control knob (F) it is possible to pass from one of the following parameters to the next one. Click the knob to modify:

 

 

 
Click the stirring control knob (F) to display the set value. Turn the knob to set the new value and click to confirm.  
To leave settings mode wait 10 seconds without touching any knob or click the lock key. 

6.  Error messages 

When the display shows an error message, the stirring function stops automatically.  
 

Error code 

Cause 

AL1 

Motor doesn’t start stirring 

AL2 

High internal motor temperature 

AL3 

Motor overload  

AL4 

High driver temperature 

Parameter shown 

on display 

Default value 

Range 

Description 

Mot 

Mot 1 

Mot 1 
Mot 2 
Mot 3 

Type motor: 

  Mot 1:

 

A00000415 

  Mot 2: A00000416 

  Mot 3: A00000417 

NOdE 

StOP 

StOP - run 

Set starting mode: 

  Stop: when the instrument is switched on through 

the main switch, OFF is on display 

  Run:  when  the  instrument  is  switched  on,  it 

restarts to work with the last set points set 

LINt 

500 

100 

 500 

Set full scale speed: 

  It limits the maximum value of speed set point 

(from 100rpm, step 100rpm) 

COun 

 

 

Running time. Days (24H): 

  Motor  working  time  is  displayed  in  hours  until 

9999 hours (around 416 days).  
After that, 

it’s shown in days starting from 417 and 

a  dot  turns  on  as  following,  to  differentiate  days 
from hours “.417” 

UPGr 

 

 

Software upgrading: 

  It allows to install a new software version 

rSEt 

nO 

YES - nO

 

Reset: 

  It allows to restore default setting parameters 

Summary of Contents for F201A0505

Page 1: ...solo per applicazioni di laboratorio per uso interno La societ produttrice declina ogni responsabilit sull impiego non conforme alle istruzioni degli strumenti Se il prodotto viene utilizzato in un mo...

Page 2: ...k muss gleich sein Fasten the unit to the support rod A00001300 using the fixing block 10007953 Position the instrument with a distance from the wall of 30 cm at least Fissare saldamente lo strumento...

Page 3: ...in einem Wasserbad oder zum Umr hren von sch dlichen Fl ssigkeiten verwendet werden bevor Schutzma nahmen gem den Sicherheitsnormen der verarbeiteten Produkte durchgef hrt werden und oder in den Labo...

Page 4: ...LI DI FUNZIONAMENTO 10 4 CONNESSIONI ESTERNE 11 5 MEN 11 6 MESSAGGI DI ERRORE 12 7 MANUTENZIONE E PULIZIA 12 8 CARATTERISTICHE TECNICHE 13 9 ACCESSORI 13 10 INFORMAZIONI MOTORI 13 1 15 2 16 3 16 4 17...

Page 5: ...is reversible Circular connectors Model MIL C 505 MS3102A 14S 6 Circular connectors Model MIL C 505 MS3102A 14S 2 A 7 Segment Display LED B Torque indicator bar C Clockwise rotation LED D Reverse key...

Page 6: ...3 Working Commissioning Switch on the instrument through the main switch J Display A shows software version the last set point value and OFF if Mode is set on Stop see chpt 5 Stirring Adjust speed set...

Page 7: ...wait 10 seconds without touching any knob or click the lock key 6 Error messages When the display shows an error message the stirring function stops automatically Error code Cause AL1 Motor doesn t s...

Page 8: ...Working hours 9 Accessories A00000415 Motor 10 lb in 417 rpm 1 11 HP 24V A00000416 Motor 28 lb in 500 rpm 1 4 HP 24V A00000417 Motor 55 lb in 250 rpm 1 4 HP 24V 10 Motor information Code A00000415 A00...

Page 9: ...9 A00000415 A00000416 A00000417...

Page 10: ...llegare lo strumento ad un motore 4 poli 24 V HP Max 8 3 A Figura 1 Vista lato connettori Assicurarsi che il valore di tensione di alimentazione dello strumento corrisponda al valore di tensione di re...

Page 11: ...senso antiorario E Coppia La barra indicatrice di coppia B costituita da 4 LED mostra un idea dell intensit di coppia applicata Il CONTROLLER Digital non mostra un valore di coppia preciso come per l...

Page 12: ...Tempo di funzionamento Giorni 24H Il tempo di lavoro del motore riportato in ore fino a 9999 circa 416 giorni Successivamente vengono mostrati i giorni a partire da 417 Un punto compare come di segui...

Page 13: ...funzionamento motore Coppia Massima coppia ammessa 280 Ncm 8 3 A Contatore Contatore motore Ore di funzionamento 9 Accessori A00000415 Motore 10 lb in 417 rpm 1 11 HP 24V A00000416 Motore 28 lb in 50...

Page 14: ...14 A00000416 A00000417...

Page 15: ...15 1 CONTROLLER Digital 100L MIL C 505 MS3102A 14S 6 MIL C 505 MS3102A 14S 2 A 7 LED B C LED D E LED F G LED H I LED J K USB L M 6P N 4P CONTROLLER Digital C E G H I J A B D F K K L K M K N K...

Page 16: ...16 2 CONTROLLER Digital 4 24 V HP 8 3 A 2 T 70 C C14 3 J A OFF 5 F H HH MM...

Page 17: ...C E B LED CONTROLLER Digital Advance H 3 H 3 LED I 4 USB USB PC 5 F 3 F Mot Mot 1 Mot 1 Mot 2 Mot 3 Mot 1 A00000415 Mot 2 A00000416 Mot 3 A00000417 NOdE StOP StOP run OFF LINt 500 100 500 100rpm 100rp...

Page 18: ...18 F 10 6 AL1 AL2 AL3 AL4 AL5 VELP Scientifica service velp it 7 UPGr rSEt nO YES nO VELP...

Page 19: ...19 8 F201A0505 115 230V 10 50 60 Hz x x 90x180x170 3 54x7 10x6 70 in 1 5 kg 3 30 lb 300 W 60 dBa 5 40 C 10 60 C 80 II CEI EN61010 1 2 2000 m 35 500 rpm 1 280 Ncm 8 3 A...

Page 20: ...HP 24V 10 A00000415 A00000416 A00000417 Motore 10 lb in 417 rpm 1 11 HP 24V Motore 28 lb in 500 rpm 1 4 HP 24V Motore 55 lb in 250 rpm 1 4 HP 24V N cm 113 316 621 lb in 10 28 55 rpm 417 500 250 V 24...

Page 21: ...ctrico Elektromotor 4 USB board Scheda USB Carte USB Tarjeta USB USBkarte USB 5 Display board Scheda display Carte display Tarjeta display Displaykarte 6 6 poles connector Connettore 6 poli Connecteu...

Page 22: ...cal equipment for measurement control and laboratory use Part 2 051 Particular requirements for laboratory equipment for mixing and stirring EN 61326 1 Electrical equipment for measurement control and...

Page 23: ...measurement control and laboratory use Part 1 General requirements EN 61010 2 051 Particular requirements for laboratory equipment for mixing and stirring EN 61326 1 Electrical equipment for measurem...

Page 24: ...zers Digestion Units Distillation Units Solvent Extractors Fiber Analyzers Dietary Fiber Analyzers Oxidation Stability Reactor Consumables Analytical Instruments Analizzatori Elementari Digestori e Mi...

Reviews: