15
Start Time
➢
It allows to set an hour at which the method starts.
➢
Enter in the menu, rotate speed control knob (C) till the desired hour. Click the knob to confirm.
➢
Even if a Start Time is set, the method begins if START button in method list is clicked.
Intermittent
Mode
➢
It allows to set stirring period and stop period alternatively.
➢
Enter in the menu, rotate speed control knob (C) till the desired working time. Click to confirm.
➢
Rotate speed control knob (C) till the desired pause time. Click to confirm.
➢
Click again the knob to modify set values.
Loop
➢
It allows to repeat the whole method N times.
➢
Enter in the menu, rotate speed control knob (C) till the desired value. Click to confirm.
Click RESET to erase all method values.
7.3 Graph
In this submenu is possible to visualize the graph of the current analysis.
Click RESET to erase the graph. Only the last 60 minutes are shown. Click ZOOM to see the last 3 minutes.
When a method is set graph appears as in
. When a method is not set, graph appears as in
7.4 Safety
In this submenu
it’s possible to set all limits linked to the safety depending on the working conditions
Speed Limit
➢
It allows to set the speed full scale.
➢
The maximum speed value is set by default.
➢
Enter in the menu, rotate speed control knob (C) by 100 rpm steps till the desired value. Click to confirm.
Torque Limit
➢
It allows to set the torque limit.
➢
The maximum value is set by default.
➢
Enter in the menu, rotate speed control knob (C) by 10 Ncm steps till the desired value. Click to confirm.
Temperature
Limit
➢
It allows to set a speed reduction if Pt100 detects a certain temperature increase (or decrease) in a defined
time.
Figure 13.
Graph with method
Figure 14.