background image

Velp Scientifica 

– ECO8 Manual

 

 

10001328/A1

 

1. Introduction 

 

The thermoeactor ECO 8 has a capacity of 8 round glass test tubes with 16 
mm external diameter and for 1 test tube with 22 mm external diameter. It is 
devised for sample preparation in determining COD, total phosphorus, total 
chromium and other parameters in water and sludge.  
This instrument, together with photometers, reagents and accessories for 
colorimetric analysis, forms a complete analytical set capable of high 
precision and reproducibility, with a limited space requirement. 
 

 
 
 

2. Instrument description 

 

The metal structure covered by an epoxy paint, purposely studied and tested, 
makes the instrument highly resistant to the chemical and mechanical 
corrosion. The block holding the test tubes is manufactured from an 
aluminium alloy which allows a very good thermal homogeneity at all the 
selectable temperatures. The temperature of the heating block is controlled by 
a microprocessor with a PID logic. The temperature probe is a Pt100 probe 
which does not require any calibration because the electronics after each 
turning on performs a self calibration. All this allows a very good precision and 
reproducibility. The ECO8 thermoreactor is equipped by a voltage changer, 
accessibile on the rear of the instrument, allowing a connection to electric nets 
230 V or 115 V. 
Four different work temperatures can be selected: 70 

– 100 – 120 – 150°C, 

with four different times of permanence of the temperature of the heating 
block at the selected value: 30 

– 60 – 120 minutes or infinite time. 

 
The unit is equipped by safety devices assuring high protection in all the 
phases of use: a cover manufactured from Teflon and a transparent cap to be 
placed over the test tubes during heating. 
 
The heating block is equipped by a protection against over temperatures 
given by a thermostat which turns off the heating when the temperature 
exceeds 180°C. 

 
 

 

3. Instrument Accessories 

The instrument is equipped with the following: 

 

Cable for electrical connection to a European plug 

 

Cable for electrical connection to an American plug 

 

Summary of Contents for ECO 8

Page 1: ...be compatible with the temperature set on the unit 3 The instrument is equipped by a voltage changer accessibile on the rear panel which allows the connection to electric nets with 230 V or 115 V Bef...

Page 2: ...and installation 4 5 Operation 4 5 1 Description of controls 4 5 2 Power interruptions 5 5 3 Safety features 5 6 Start of a cycle 6 7 Maintenance 6 7 1 Replacing fuses 6 8 Disposal 7 9 Accessories 7...

Page 3: ...a microprocessor with a PID logic The temperature probe is a Pt100 probe which does not require any calibration because the electronics after each turning on performs a self calibration All this allow...

Page 4: ...he key lower the programmed temperature to 150 120 100 or 70 C The selected temperature is shown by the lighting of the corresponding led It is possible to select different temperatures if the program...

Page 5: ...t The LED blinks to signal that the preselected temperature has been reached and the time count has started 5 2 Power Interruptions The instrument will register a black out or a lack of power supply a...

Page 6: ...will sound continuously for 5 seconds At this time the heating element and the LED s corresponding to time and temperature will turn off The instrument is now ready for a new work cycle If the selecte...

Page 7: ...ture oscillations around the set point The Pt100 probe provides high precision and does not require calibration because the internal software performs a self calibration every time the instrument is t...

Page 8: ...22 x n 1 Thermoregulation P I D microprocessor Precision 0 2 end of scale Temperature probe Pt 100 class A Probe calibration Automatic by software Heating block Temperature stability C 0 3 Temperature...

Page 9: ...fica ECO8 Manual 10001328 A1 9 12 Electric scheme 1 5x20mm retarded fuse 3 15A 2 General lighted switch 3 Voltage changer electronic card 4 Safety thermostat 5 Electric resistance 6 Electronic card 7...

Page 10: ...Velp Scientifica ECO8 Manual 10001328 A1 10 13 Suggestions Any suggests and advice would be greatly appreciated to enable us to constantly improve this instruction manual e mail service velp it...

Page 11: ...t the product is in according with the following standards EN 61010 1 2001 EN 61326 1 2000 And satisfies the essential requirements of directives Machines directive 98 37 CEE Low voltage directive 73...

Page 12: ...ur product Lines Environment C O D B O D B O D Sensor REFRIGERATED THERMOSTATS INCUBATORS JAR TEST METALS IN SLUDGES AND SEDIMENTS NEPHELOMETRIC TURBIDITY NITROGEN IN DIFFERENT FORMS PERFORMANCE OF AN...

Reviews: