When using the measuring probe in the X10
position, the measuring impedance will
increase to 10Mohm, thereby reducing the
measuring instrument's charge on the
coupling.
Press the “x1/x10” key in order to adjust the
display to the setup of the measuring probe.
Upon choosing X10, “X10” will be indicated
on the screen and all measuring values will
be adjusted accordingly.
IMPORTANT : X10 measuring probes should be calibrated. Read
“Setting up a X10 measuring probe".
Auto-setup function
The auto-setup function is ideally suited for quick measurements as no
manual setup has to be made and everything is automatic.
F
Use the auto-setup function when the screen no longer displays a
signal after the manual setup.
•
Connect the signal to be checked to the BNC input by means of a
special measuring lead or measuring probe.
•
Activate the unit.
•
Press “Auto” in order to (de)activate the auto-setup function.
When “auto setup” is engaged, the Volt/div and Time/div readouts are
highlighted.The auto setup function will provide correct measurements
for repetitive and stable signals of max. 400KHz.
The display may read as follows :
REMARK:
The y-position does not work in auto-setup, which is why the trigger
setup remains in run or normal mode.The auto-setup function will
deactivate when using the Trigger , Volt and time per division and Y
position functions. (see below).
x1 x10
x1 x10
Auto