To replace the battery or fuse:
Switch of the meter
Remove the protective edge, release the two screws located at the
backside of the meter and gently open the housing.
Remove the old battery and insert a new one.
Battery
: 9V 6LR61 alkaline, do not use rechargeable batteries and
make sure to respect the polarity
Fuses
: F500mA 600V and F10A 600V 5x20mm
Make sure the meter is closed tight and put the protective edge back
in place before using the meter.
Troubleshooting
If the device beeps continuously while measuring continuity, this
means that the 600V-500mA internal fuse is defective. Replace this
fuse.
Keep in mind that a low battery level could lead to incorrect
measurements. Replace the 9V battery on a regular basis.
(tip: the reduced luminosity of the backlight/LCD display indicates a
low battery level)