9.
Technische specificaties
This device is not calibrated when purchased!
Regulations concerning environment of use:
•
Use this meter only for measurements in CATI, CATII, CAT III (<1000V) and CAT
IV (<600V) environments (see §4)
•
Use this meter only in a pollution degree 2 environment (see §5)
Ideal working conditions include:
•
Temperature: 18 to 28°C (64°F to 82°F)
•
relative humidity: max. 75%
•
altitude: max. 2000m (6560ft)
9.1.
General
Max. Voltage
600V CAT IV, 1000V CAT III
Display
3
5/6-
digit LCD, 2 to 3 samples/sec.
Fuse Protection
µA/mA-range: 630mA/1000V (Ø10.3x38mm)
10A-range: 10A/1000V (Ø10.3x38mm)
Temperature coefficient
0.1x(specified accuracy)/°C (<18°C or >28°C)
Power Supply
9V alkaline battery 6LR61
Ranging Method
manual/automatic
Polarity Indication
" - " is displayed
Overrange Indication
"OL" is displayed automatically
Battery-Low Indication
" " is displayed
Operating Temperature
0°C to 40°C (32~122°F) (<80%RH, <10°C non
condensing)
Storage Temperature
-10°C to 60°C (14~140°F) (<70%RH, battery removed)
Dimensions
190x90x40mm
Weight
±500g (incl. battery)
9.2.
DC Voltage
Range
Resolution
Accuracy
600mV
100µV
±
(0.5% of rdg + 5 digits)
6V
1mV
±
(0.8% of rdg + 5 digits)
60V
10mV
600V
100mV
1000V
1V
±
(1.0% of rdg +2 digits)
Input Impedance: >10M
Ω
for all ranges
9.3.
AC Voltage
Range
Resolution
Accuracy
600mV
100µV
±
(1.0% of rdg + 5 digits)
6V
1mV
60V
10mV
600V
100mV
1000V
1V
±
(1.5% of rdg +2 digits)
Input Impedance: >10M
Ω
for all ranges
Frequency Range: 40 to 400Hz
Overload protection: 1000Vdc or ac rms.
9.4.
DC Current
Range
Resolution
Accuracy
600µA
0.1µA
±
(1.5% of rdg + 3 digits)
6mA
1µA
60mA
10µA
600mA
0.1mA
6A
1mA
±
(1.5% of rdg +5 digits)