1.1. SAFETY

This chair has CE certification and ap-
proval for the model name “VELA Rumba” 
and adheres to EU regulations. The chair 
adheres to all relevant Electromagnetic 
Compatibility (EMC) requirements and EU 
requirements for electrical safety.

The chair is manufactured by:
VELA Medical, Gøteborgvej 8-12, 
9200 Aalborg SV, Denmark.

1.1.1. SAFETY NOTES
::

  The chair is intended for indoor use in 

hospital settings.

::

  The chair is not suitable for use outside 

hospital settings.

::

  The chair does not have electrical ac-

cessories.

::

  The chair must be kept at least 30 cm 

from other electrical equipment, includ-
ing electrical wires/cables.

::

  The chair must not be used for opera-

tions or in contagious wards, as the chair 
cannot be run through a washer.

::

  We recommend annual servicing where 

the chair is checked for defective or 
worn elements.

::

  Batteries must be disposed of at recy-

cling plants in the respective containers 
for batteries/electrical garbage (refer to 
section 1.5 Recycling and Disposal). 

::

  The chair must be cleaned and main-

tained in accordance with Section 1.4 
Maintenance.

::

  The footrest is not designed to be sat 

upon.

::

  The armrest is not designed to be sat 

upon.

::

  The backrest is not designed to be sat 

upon when lowered to a position where 
this is possible.

::

  When adjusted to the Trendelenburg 

position, the chair has been designed to 
take the patient’s weight only.

::

  The maximum load for the chair is 140 kg.

::

  Take care not to get hands or limbs 

trapped in moving parts when using the 
electrical features (marked with yellow 
warning labels).

::

  The chair should not be sat upon during 

transportation.

::

  To prevent unintended adjustments, all 

electrical features are inactivated when 
the remote control button is released.

::

  With annual servicing, the lifespan of the 

chair is 5 years of daily use up to 8 hours 
a day. 

::

  The chair is only intended for use by 

adults.

::

  Make sure that cables and service cables 

do not pose a tripping hazard.

::

  Make sure the chair is not driven over 

cables or service cables.

If the chair runs on battery, make sure the 
battery is charged 24 hours before use!

GENERAL

Summary of Contents for Treatment 87213

Page 1: ...VELA Medical U S E R M A N U A L E N Manual nr 105634 Item no Model 87211 VELA Rumba 87212 VELA Neonatal chair 87213 VELA Treatment Chair The chair s design varies depending on the model...

Page 2: ...Seat angle and footrest 8 1 7 3 Backrest angle 8 1 7 4 Neutral and Trendelenburg positions 8 1 8 SETTINGS 9 1 8 1 Using the brakes 9 1 8 2 Using the footrest 9 1 8 3 Adjusting the armrest 9 1 8 4 Adj...

Page 3: ...r please read this user guide carefully and keep it handy for future reference You can also find this user guide on VELA Medical s website www vela medical com along with other relevant information ab...

Page 4: ...ge refer to section 1 5 Recycling and Disposal The chair must be cleaned and main tained in accordance with Section 1 4 Maintenance The footrest is not designed to be sat upon The armrest is not desig...

Page 5: ...ransport specifications Temperature 40 C to 70 C refer to Section 1 4 2 Upholstering Humidity 15 to 95 Air pressure PPP up to 1 000 metres above sea level 1 2 WARRANTY VELA Medical offers warranties i...

Page 6: ...d 70 Ethanol hospital disinfectant ISO 1164 or chlorine based products with 5 000 ppm active chlorine such as Actichlor Plus We advise following the instructions on the packaging After cleaning disinf...

Page 7: ...nst the wooden frame the length compensation is working as intended and the backrest pillow has been mounted correctly When removing pillows follow the instructions in reverse 1 6 2 SEAT 1 Place seat...

Page 8: ...3b to angle the backrest backwards Warning Do not sit on the backrest 1 7 1 LIFT SEAT HEIGHT Press the button on the left 1a to lower the chair or press the button on the right 1b to raise the chair W...

Page 9: ...t 1 8 2 USING THE FOOTREST The footrest follows the seat angle Refer to Section 1 7 2 Seat Angle and Footrest Warning Do not sit on the footrest 1 8 1 USING THE BRAKES Press down on the brake pedal on...

Page 10: ...turns green once the battery is fully charged The LED light can only be seen when alighting assistance is activated The chair should be charged every day Not all chairs are equipped with a battery 1...

Page 11: ...Figure 2 Note footrest follows the seat angle Warning Beware of trapping hands or limbs underneath the chair when using alighting assistance 1 9 1 LIFT UP OR DOWN The button on the left lowers the se...

Page 12: ...Seat down Note if the chair is equipped with the Trendelenburg feature this feature also resets to the neutral position Section 1 9 6 1 9 3 BACKREST ANGLE The button on the left angles the backrest fo...

Page 13: ...eutral 2 Backrest up and 3 Seat down Note not all chairs are equipped with the Trendelenburg feature 1 9 5 TRENDELENBURG The red button on the right angles the seating unit to Trende lenburg backrest...

Page 14: ...ical potential 230V for regular sockets Battery additional purchase Li Ion 25 2V 1800 mAH IP66 2 0 TECHNICAL DATA AND DOCUMENTATION STORAGE AND TRANSPORTATION The chair is intended for indoor use and...

Page 15: ...C 61000 4 2 8 kV contact disharge 2 4 6 8 15 kV air discharge Radiated RF field IEC 61000 4 3 3 V m 80 MHz 2 7 GHz 80 AM 1 kHz Proximity fields from wireless transmitters IEC 61000 4 3 80 MHz to 2700H...

Page 16: ...701 551 4 Vermund Larsen A S VELA Medical G teborgvej 8 12 9200 Aalborg SV Denmark Tel 45 96 34 76 00 mail vela medical com vela medical com...

Reviews: