etup procedure
A
fter
VEGASON
63
is connected to voltage supply or after
voltage recurrence
,
the instrument carries out a self
-
check for
approx
.
30
seconds
.
T
he following steps are carried out
:
l
I
nternal check of the electronics
l
I
ndication of the instrument type
,
the
fi
rmware as well as
the sensor
TAG
s
(
sensor designation
)
l
O
utput signal jumps brie
fl
y
(
approx
.
10
seconds
)
to the set
fault current
T
hen the actual measured value is displayed and the
corresponding current is transmitted to the cable
.
2
)
I
n
HART
-
M
ultidrop mode
(
several sensors on one input
)
the
address must be set before continuing with the parameter
adjustment
.
Y
ou will
fi
nd a detailed description in the operating
instructions manual
"
I
ndicating and adjustment module
"
or in
the online help of
PACT
ware
™
or
DTM
.
HART
mode
S
tandard
A
ddress
0
A
s
VEGASON
63
is a distance measuring instrument
,
the
distance from the sensor to the product surface is measured
.
T
o have the real product level displayed
,
an allocation of the
measured distance to the percentage height must be made
.
T
o
carry out this adjustment
,
the distance is entered with full and
empty vessel
.
I
f these values are not known
,
an adjustment
with the distance values
,
e
.
g
.
10
%
and
90
%
is also possible
.
S
tarting point for these distance speci
fi
cations is always the
the lower side of the
fl
ange
(
fl
ange version
,
for all other
versions the lower side of the transducer
.
T
he actual level is then calculated on the basis of these
entered values
.
A
t the same time
,
the operating range of the
sensor is limited from maximum range to the requested range
.
T
he real product level during this adjustment is not important
,
because the min
./
max
.
adjustment is always carried out
without changing the product level
.
T
hese settings can be
made ahead of time without the instrument having to be
installed
.
2
)
T
he values correspond to the actual measured level as well as to the set
-
tings already carried out
,
e
.
g
.
default setting
.
S
witch
-
on phase
A
ddress setting
HART
-
M
ulti
-
drop
P
arameter adjustment
S
et up with the indicating and adjustment module
PLICSCOM
VEGASON
63
-
4
...
20
m
A
/
HART
31
28777
-