12
4 Mounting
VEGAPULS 61 • Profibus PA
28444-EN-141215
4 Mounting
4.1 General instructions
Select an installation position you can easily reach for mounting and
connecting as well as later retrofitting of a display and adjustment
module. The housing can be rotated by 330° without the use of any
tools. You can also install the display and adjustment module in four
different positions (each displaced by 90°).
Warning:
With threaded versions, the housing must not be used to screw in the
instrument! Applying tightening forces on the housing can damage its
internal parts.
Use the recommended cables (see chapter "
Connecting to power
supply
") and tighten the cable gland.
You can give your instrument additional protection against moisture
penetration by leading the connection cable downward in front of the
cable entry. Rain and condensation water can thus drain off. This ap-
plies mainly to outdoor mounting as well as installation in areas where
high humidity is expected (e.g. through cleaning processes) or on
cooled or heated vessels.
Fig. 3: Measures against moisture ingress
The reference plane for the measuring range of the sensor depends
on the antenna version.
With
encapsulated antenna system
, the reference plane is the seal
surface of the thread.
Installation position
Screwing in
Moisture