background image

9

VEGAPULS 61

37991-EN-200303

such as e.g. plastic housing, metal housing with inspection window or plastic antenna, have a cau-

tion label pointing out the safety measures that must be taken with regard to electrostatic charges 

during operation.

WARNING -  POTENTIAL  ELECTROSTATIC

CHARGING HAZARD  -  SEE INSTRUCTIONS

Caution: Plastic parts! Danger of electrostatic charging!

• 

Avoid friction

• 

No dry cleaning

• 

Construction/Installation: The VEGAPULS PS61(*).C****P/K/F/L**** must be constructed/

installed in such a way that

 

– electrostatic charges are ruled out during operation, maintenance and cleaning.

 

process-related electrostatic charges, e.g. by measuring media flowing past, are ruled out

Use of an overvoltage arrester

If necessary, a suitable overvoltage arrester can be connected in front of the VEGAPULS 

PS61(*).C****P/K/F/L****.
When used as category 1G or 1/2G instrument, a suitable overvoltage arrester must be connected 

in between as protection against overvoltages, as far as required according to EN 60079-14.

Grounding

In order to avoid the danger of electrostatic charging of the metallic parts, the VEGAPULS 

PS61(*).C****P/K/F/L**** must be electrostatically connected to the local potential equalisation 

(transfer resistance ≤ 1 MΩ), e.g. via the ground terminal, when used as category 1G or 1/2G instru

-

ments.

Metallic adapter flanges must be earthed, especially when installed on non-conductive plastic ves

-

sels or non-earthed vessels. Earthing can be realised with cable lug via the flange gland.

Impact and friction sparks

When used as category 1G instruments, the VEGAPULS PS61(*).C****P/K/F/L**** in aluminium/

titanium versions must be mounted in such a way that sparks from impact and friction between alu-

minium/titanium and steel (except stainless steel, if the presence of rust particles can be excluded) 

cannot occur. 

10  Material resistance

For applications requiring instruments of category 1G or category 1/2G the VEGAPULS 

PS61(*).C****P/K/F/L**** must only be used in products against which the wetted materials are suf

-

ficiently resistant.

11  Mounting with external display unit VEGADIS 61/81

The intrinsically safe signal circuit between VEGAPULS PS61(*).C****P/K/F/L**** and the external 

indicating unit VEGADIS 61/81 should be set up without grounding. The required insulation voltage 

is > 500 V AC. When using the VEGA connection cable included with the delivery, this requirement 

is fulfilled. If grounding of the cable screen is required, it must be carried out according to EN 60079-

14.

Summary of Contents for VEGAPULS 61 C Series

Page 1: ...8 7 Use of an overvoltage arrester 9 8 Grounding 9 9 Impact and friction sparks 9 10 Material resistance 9 11 Mounting with external display unit VEGADIS 61 81 9 12 Removing and replacing the red thre...

Page 2: ...k ytt varten EL DE Die vorliegenden Sicherheitshinweise sind im Download unter www vega com standard m ig in den Sprachen deutsch englisch franz sisch und spanisch verf gbar Weitere EU Landessprachen...

Page 3: ...tible liquids gases mist or vapour The VEGAPULS PS61 C P K F L are suitable for use in hazardous atmospheres of all combustible materials of explosion group IIA IIB and IIC for applications requiring...

Page 4: ...s L Two wire Foundation Fieldbus with increased sensitivity g Housing Protection K Plastic single chamber IP66 IP67 A Aluminium single chamber IP66 IP68 0 2 bar H Special colour Aluminium single chamb...

Page 5: ...ns then these will be mentioned explic itly with their type code 4 Technical data Electrical data Type of protection intrinsic safety Ex i Power supply and signal circuit termi nals 1 2 in electronics...

Page 6: ...insic safety Ex ia IIC Only for connection to the display and adjust ment module PLICSCOM or VEGACONNECT PTB 07 ATEX 2013 X With the double chamber housing version the display and adjustment module ma...

Page 7: ...or also in case of failure does not generate heat itself Responsibility for safe operation of the equipment with respect to pressures temperatures of the materials used rests with the operator The pre...

Page 8: ...trostatically connected to the local potential equalisation transfer resistance 1 M e g via the ground terminal when used as category 1G or 1 2G instru ments Metallic adapter flanges must be earthed e...

Page 9: ...ice and the openings must be sealed according to the requirements of the type of protection and the IP protection type specified on the type label When using certified i e suitable cable glands sealin...

Page 10: ...11 Notes VEGAPULS 61 37991 EN 200303...