background image

5

VEGAPULS 61

37991-EN-200303

Important specification in the type code

VEGAPULS PS61(*).abcdefghij

Position

Feature

Description

ab Approval

CX

ATEX II 1G, 1/2G, 2G Ex ia IIC T6

CA

ATEX II 1G, 1/2G, 2G Ex ia IIC T6 + WHG

CM

ATEX II 1G, 1/2G, 2G Ex ia IIC T6 + Ship approval

c

Version / Material / Pro

-

cess temperature

A

with encapsulated horn antenna (ø 40 mm) / PVDF / -40 … +80 °C

B

with plastic horn antenna (ø 80 mm) / PP / -40 … +80 °C

de

Process fitting / Material

**

Threaded connection, Clamp, flanges; two-digit alphanumeric code 

for metallic process fittings, industrial flanges according to ASME, 

BS, DIN, EN, GOST, HG/T, JIS and for other international, national or 

industrial standards, guidelines or standards with suitable pressure 

and temperature specifications

f

Electronics

P

Two-wire Profibus PA

K

Two-wire Profibus PA with increased sensitivity

F

Two-wire Foundation Fieldbus

L

Two-wire Foundation Fieldbus with increased sensitivity

g

Housing / Protection

K

Plastic single chamber / IP66/IP67

A

Aluminium single chamber / IP66/IP68 (0.2 bar)

H

Special colour Aluminium single chamber / IP66/IP68 (0.2 bar)

3

Aluminium single chamber / IP66/IP68 (1 bar)

D

Aluminium double chamber / IP66/IP68 (0.2 bar)

S

Special colour Aluminium double chamber / IP66/IP68 (0.2 bar)

Y

Aluminium double chamber / IP66/IP67 with M12 x 1 for VEGADIS 

61/81

V

Stainless steel single chamber (precision casting) / IP66/IP68 

(0.2 bar)

5

Stainless steel single chamber (precision casting) / IP66/IP68 (1 bar)

8

Stainless steel single chamber (electropolished) / IP66/IP68 (0.2 bar)

W

Stainless steel double chamber / IP66/IP68 (0.2 bar)

Q

Stainless steel double chamber / IP66/IP67 with M12 x 1 for 

VEGADIS 61/81

R

Plastic double chamber / IP66/IP67

X

Plastic double chamber / IP66/IP67 with M12 x 1 for VEGADIS 61/81

h

Cable entry / Cable 

gland, Plug connection

M

M20 x 1.5 / without

N

½ NPT / without

*

One-digit alphanumerical code for further suitable fittings, cable en

-

tries and closing screws.

Summary of Contents for VEGAPULS 61 C Series

Page 1: ...8 7 Use of an overvoltage arrester 9 8 Grounding 9 9 Impact and friction sparks 9 10 Material resistance 9 11 Mounting with external display unit VEGADIS 61 81 9 12 Removing and replacing the red thre...

Page 2: ...k ytt varten EL DE Die vorliegenden Sicherheitshinweise sind im Download unter www vega com standard m ig in den Sprachen deutsch englisch franz sisch und spanisch verf gbar Weitere EU Landessprachen...

Page 3: ...tible liquids gases mist or vapour The VEGAPULS PS61 C P K F L are suitable for use in hazardous atmospheres of all combustible materials of explosion group IIA IIB and IIC for applications requiring...

Page 4: ...s L Two wire Foundation Fieldbus with increased sensitivity g Housing Protection K Plastic single chamber IP66 IP67 A Aluminium single chamber IP66 IP68 0 2 bar H Special colour Aluminium single chamb...

Page 5: ...ns then these will be mentioned explic itly with their type code 4 Technical data Electrical data Type of protection intrinsic safety Ex i Power supply and signal circuit termi nals 1 2 in electronics...

Page 6: ...insic safety Ex ia IIC Only for connection to the display and adjust ment module PLICSCOM or VEGACONNECT PTB 07 ATEX 2013 X With the double chamber housing version the display and adjustment module ma...

Page 7: ...or also in case of failure does not generate heat itself Responsibility for safe operation of the equipment with respect to pressures temperatures of the materials used rests with the operator The pre...

Page 8: ...trostatically connected to the local potential equalisation transfer resistance 1 M e g via the ground terminal when used as category 1G or 1 2G instru ments Metallic adapter flanges must be earthed e...

Page 9: ...ice and the openings must be sealed according to the requirements of the type of protection and the IP protection type specified on the type label When using certified i e suitable cable glands sealin...

Page 10: ...11 Notes VEGAPULS 61 37991 EN 200303...