10 Diagnostics and servicing
VEGAPOINT 23 • Transistor (PNP/NPN)
56627-EN-210210
Code
Text message
Cause
Rectification
S604
Switching output overloaded
Overload or short circuit on the switch-
ing output
Check electrical connection
Code
Text message
Cause
Rectification
M511
Inconsistent software con-
figuration
Software faulty or outdated
Carry out software update
10.5 Software update
The device software is updated via Bluetooth.
The following components are required:
•
Instrument
•
Voltage supply
•
PC/notebook with PACTware/DTM and Bluetooth USB adapter
•
Current instrument software as file
You can find the current instrument software as well as detailed infor-
mation on the procedure in the download area of our homepage.
Caution:
Instruments with approvals can be bound to certain software versions.
Therefore make sure that the approval is still effective after a software
update is carried out.
You can find detailed information in the download area on our home-
page.
10.6 How to proceed if a repair is necessary
You can find an instrument return form as well as detailed information
about the procedure in the download area of our homepage. By doing
this you help us carry out the repair quickly and without having to call
back for needed information.
In case of repair, proceed as follows:
•
Print and fill out one form per instrument
•
Clean the instrument and pack it damage-proof
•
Attach the completed form and, if need be, also a safety data
sheet outside on the packaging
•
Ask the agency serving you to get the address for the return ship-
ment. You can find the agency on our homepage.