5
2 Mounting
Screening tube adapter • for VEGACAL 63, 64/VEGACAP 63
34849-EN-130916
Caution:
Check the insulation of the probe for deep scratches.
To ensure the seal effect of the screening tube adapter, the isolation
of the probe must be undamaged in the area of the lower screening
tube adapter end.
1.
Loosen and remove (SW
24) pressure screw (5)
2.
Remove the two O-rings (4) from the screening tube
3.
Place the process seal (2) on the thread of the probe
4.
Push the rod of the probe (1) through the screening tube adapter
(3)
5.
Screw the screening tube adapter (3) onto the thread of the probe
and tighten with 100
Nm (74
lbf
ft) (SW
46)
6.
If it is harmless to the process, grease the two O-rings with an
acid-free grease so that the rod of the probe slides more easily
7.
Slide the two O-rings (4) over the rod of the probe
8.
Slide the pressure screw (5) of the screening tube adapter onto
the probe
9.
Screw pressure screw (5) with the two O-rings into the screening
tube adapter and tighten it to the stop (SW
24)
The mounting of the screening tube adapter is now finished. The
probe can now be installed.
You can find the mounting procedure in the operating instructions of
the probe.
Note:
The VEGACAP probe must be re-compensated after mounting the
screening tube adapter. Use the compensation key on the electronics
module. You can find the instructions for this in the operating instruc
-
tions of the probe.