55
10 Supplement
VEGABAR 66 • 4 … 20 mA/HART - climate compensated
36738-EN-130321
Ʋ
Ohmic contact
Between ground terminal, process fitting and transmitter
Ʋ
Connection cable between IP
68
housing and external electronics
PUR, FEP, PE
Ʋ
Type label support with IP
68 version
on cable
PE hard
Weight approx.
Ʋ
Basic weight
0.7
kg (1.543
lbs)
Ʋ
Suspension cable
0.1
kg/m (0.07
lbs/ft)
Ʋ
Connection tube
1.5
kg/m (1
lbs/ft)
Ʋ
Straining clamp
0.2
kg (0.441
lbs)
Ʋ Threaded fitting
0.4
kg (0.882
lbs)
Lengths
Ʋ
Connection tube
0.25
…
6
m (0.82
…
19.69
ft)
Output variable
Output signal
4
…
20
mA/HART
HART output values according to HART standard 5.0
Ʋ
Primary Value
Pressure
Ʋ
Secondary Value
Temperature
10)
Signal resolution
1.6
µA
Failure signal current output (adjustable) mA-value unchanged 20.5
mA, 22
mA, <
3.6
mA
Max. output current
22
mA
Load
see load diagram under Power supply
Met NAMUR recommendation
NE
43
Dynamic behaviour output
Run-up time approx.
10
s
90 %
100 %
10 %
t
t
T
A
t
t
S
2
1
Fig. 32: Sudden change of the process variable. t
T
: dead time; t
A
: rise time; t
S
: jump response time
1 Process variable
2 Output signal
10)
Depending on the respective instrument and the process fitting, the temperature can deviate from the real
process temperature.