6 Set up with the display and adjustment module
SOLITRAC 31 • Profibus PA
43834-EN-211203
Peak values of temperature:
Resetting of the measured min. and
max. temperatures to the actual measured value.
The following table shows the default values of the instrument. The
values apply for the application "
Limit level
". First of all you have to
select the application.
Depending on the instrument version, not all menu items may be
available or they may be differently assigned:
Menu
Menu item
Default value
Setup
Sensor address
126
Measurement loop name
Sensor
Isotope
Cs-137
Application
Limit level
Background radiation
0 ct/s
Temperature unit
° C
Adjustment mode
1-point
Adjustment (uncovered)
90000 ct/s
Delta I
90 %
AI FB1 Channel
Process value
AI FB1 scaling unit
%
AI FB1 scaling
0 % = 0 %
100 % = 100 %
AI FB1 damping
0 s
Mode - Relay
Dry run protection
Lock adjustment
Released
Display
Language
Selected language
Displayed value
Pulse rate (ct/s)
With this function you can select the mode.
The sensor offers the HART modes standard and multidrop.
The default setting is standard with address 0.
If the measured value is output via the 4 … 20 mA output, you must
not switch over to HART Multidrop.
The mode 'Standard', with fixed address 0 (factory setting), means
output of the measured value as 8/16 mA signal.
In Multidrop mode, several sensors are communicating on one two-
wire cable via the HART protocol.
In Multidrop mode, up to 15 sensors can be operated on one two-wire
cable. An address between 1 and 15 must be assigned to each sen-
sor.
HART mode