6 Set up with the display and adjustment module
SOLITRAC 31 • Profibus PA
43834-EN-211203
Switching function of the relay
Information:
10 minutes after the key was pressed for the last time, the simulation
is interrupted automatically. You can also interrupt the simulation with
the switch on the electronics module.
The sensor calculates a suitable integration time automatically.
6.5.4 Additional adjustments
In this menu item you can set the actual date, time and display format.
When a reset is carried out, all settings (with only a few exceptions)
are reset. The exceptions are: PIN, language, SIL and HART mode.
The following reset functions are available:
Basic settings:
Resetting of the parameter adjustments to default
values at the time of shipment. Order-specific settings are deleted.
Default settings:
Resetting of the parameter adjustment like under
"
Basic settings
". In addition, special parameters are reset to default
values. Order-specific settings are deleted.
Peak values of measured value:
Resetting of the parameter adjust-
ments in the menu item "
Setup
" to the default values of the respec-
tive instrument. Order-specific settings remain but are not taken over
into the current parameters.
Calculated damping
Date/Time
Reset