21
4 Setup
SHLD1 • Source holder
62092-EN-190704
4 Setup
4.1 Operation of SHLD1
Warning:
Before switching on the radiation, make sure that no persons are
inside the radiated areas (also not inside the vessel).
Radiation must only be switched on by trained personnel.
The figures in brackets refer to the following illustration.
2
1
2
1
A
B
Fig. 13: Switching on the radiation with the manual operating lever - example:
SHLD1 standard version
A Source holder switched off - operating lever in position "OFF" (2)
B Source holder switched on - operating lever in position "ON" (1)
1 Switching position "ON"
2 Switching position "OFF"
Initial situation: Source holder is in position "OFF" (2)
1. Open and remove padlock.
The radiation safety officer gets a separate notification with the
code for the padlock. Contact our responsible sales organisation.
Keep the padlock near by the source holder. Do not insert the
padlock into the opening of the "OFF" position because otherwise
the source holder cannot be switched off completely in case of
emergency.
2. Screw out safety screw (3) (screw is undetachably fixed to the
safeguard cable)
3. Turn the operating lever 90° clockwise up to the stop.
"ON" (1) will then appear in the position recess of the operating
lever.
4. Secure the operating lever in position "ON" (1).
Switching the radiation
on