l
U
ser programmable
U
ser programmable means
:
S
witching on a linearization curve
programmed via
PC
and
PACT
ware
A
linearization is necessary for all vessels in which the vessel volume
does not increase linearly with the level
-
e
.
g
.
in a horizontal cylindrical
or spherical tank
-
and the indication or output of the volume is
required
.
C
orresponding linearization curves are preprogrammed for
these vessels
.
T
hey represent the correlation between the level
percentage and vessel volume
.
T
he linearisation applies to the
measured value indication and the current output
.
B
y activating the
appropriate curve
,
the volume percentage of the vessel is displayed
correctly
.
I
f the volume should not be displayed in percent but e
.
g
.
in l
or kg
,
a scaling can also be set in the menu item
"
D
isplay
"
.
F
actory setting is linear
.
L
inearisation curve
linear
C
aution
:
N
ote the following
,
if the respective sensor is used as part of an over
fi
ll
protection system according to
WHG
:
I
f a linearisation curve is selected
,
the measuring signal is no longer
compulsorily linear proportional to the level
.
T
his must be taken into
consideration by the user
,
particularly when adjusting the switching
point on the level switch
.
I
n the menu item
"
S
ensor
-
TAG
"
you edit a
12
-
digit measurement loop
name
.
A
n unambiguous designation can hence be assigned to the
sensor
,
e
.
g
.
the measurement loop name or the tank or product
designation
.
I
n digital systems and in the documentation of larger
plants
,
a singular designation should be entered for exact identi
fi
cation
of individual measuring sites
.
T
he available digits comprise
:
l
L
etters from
A
…
Z
l
N
umbers from
0
…
9
l
S
pecial characters
+
, -, /, -
F
actory setting is
"
S
ensor
"
.
S
ensor
-
TAG
S
ensor
B
asic adjustment
/
S
en
-
sor
TAG
12
I
ndicating and adjustment module
PLICSCOM
5
S
et up
-
S
ensors up to
SW
3
.
99
36433
-
EN
-
091104