56
6 Set up with the display and adjustment module
MINITRAC 31 • 4 … 20 mA/HART - four-wire
40447-EN-130430
The Master instrument must have the function "Level". For this
purpose, select under the menu item "
Setup/Application
" the function
"Level".
Set the address setting (MGC) on the Master instrument to "0 - 0".
You have to enter the address of the Slave instrument in the list of
the Master instrument. This function is not possible in the display and
adjustment module. To do this you need PACTware with the respective
DTM.
The main menu is divided into five sections with the following func
-
tions:
Setup:
Settings, e.g. for measurement loop name, isotope, applica-
tion, background radiation, adjustment, signal output
Display:
Settings, for example language, measured value display
Diagnosis:
Information, for example, of device status, peak value,
simulation
Additional adjustments:
Instrument unit, reset, date/time, copying
function
Info:
Instrument name, hardware and software version, date of manu-
facture, instrument features
Check if the correct language is already set for the display. If not, you
can change the language in the menu item "
Display/Language
".
Start with the setup of MINITRAC 31.
In the main menu point "
Setup
", the individual submenu points should
be selected subsequently and provided with the correct parameters
to ensure the optimum adjustment of the measurement. The proce-
dure is described in the following.
Possibly keep the sequence of the menu items.
Adjustment
In this menu item you can assign an unambiguous name to the sen-
sor or the measurement loop.
This parameter is described in the operating instructions manual
"
Display and adjustment module
".
Main menu
Proceeding
Setup - Measurement
loop name