Force Imaging
Force Calibration Control Panels and Menus
212
Dimension 3100 Manual
Rev. D
13.3 Force Calibration Control Panels and Menus
The
Force
Calibration
Control
window (see
) manipulates the microscope in
Force
Calibration
mode. The parameters control the rate, start position and amplitude of the triangle
wave applied to the Z piezo. You can also adjust the
Setpoint
value of the cantilever deflection
voltage used in the feedback loop during imaging. The
Capture
button stores the force curve for
Offline
viewing. Some parameters influence the operation of the microscope during imaging; most
of the menu parameters affect only
Force Calibration
mode.
Figure 13.3a
Advanced Force Calibration Control Window (Contact Mode AFM)
There are 2 methods to enter Force Calibration Mode:
•
View
>
Force Mode
>
Calibrate
will open a simplified subset of the parameters and
commands available.
•
View
>
Force Mode
>
Advanced
includes the full set of available parameters and
commands.
Additionally:
•
View
>
Force Mode
>
Step
enters force mode with the parameter
start mode
=
step
and the piezo fully retracted.
•
View
>
Force Mode
>
Volume
enters Force Volume Mode.
Items in the
Force Calibration
Control
Window are discussed individually below.
038
Summary of Contents for Dimension 3100
Page 12: ...xii Dimension 3100 Manual Rev D ...
Page 20: ...List of Figures xx Dimension 3100 Manual Rev D ...
Page 72: ......
Page 106: ......
Page 118: ......
Page 214: ...Scanning Tunneling Microscopy STM Etching Tungsten Tips 194 Dimension 3100 Manual Rev D ...
Page 222: ......
Page 266: ......
Page 274: ......