Calibration, Maintenance, Troubleshooting and Warranty
SPM Calibration Overview
280
MultiMode SPM Instruction Manual
Rev. B
15.1.2
Calibration References
As described above, each scanner exhibits its own unique sensitivities; therefore, it is necessary to
precisely measure these sensitivities, then establish software parameters for controlling the scanner.
This task is accomplished with the use of a calibration reference (see
).
Figure 15.1c
Veeco Silicon Calibration Reference
This calibration reference consists of a silicon substrate having a regular series of pits, each 200nm
deep. Pits are spaced apart on 10µm centers. This reference is typically used to calibrate J scanners.
For calibrating E or A scanners, please refer to the table below. Other, similar surfaces are available
with different dimensions. Atomic-scale calibrations are generally carried out with mica or
graphite, which exhibit very regular atomic lattices. Calibration references serve as the primary tool
by which SPMs are calibrated. They serve as measuring sticks with which to gauge scanner
displacement for a given voltage.
The SPM should be capable of measuring a calibration reference with an accuracy of 2 percent or
better while scanning at the maximum
Scan size
setting. By using
fi
ne calibration techniques, it is
possible to calibrate the SPM with greater accuracy.
Table 15.1b
Recommended Calibration Model # Associated with Specific Scanners
Scanner
HOPG
(STM)
MICA
(AFM)
498-000-026
498-000-025
PG
ISGS
Calibration
J or JV
X-Y
√
Z
√
E or
EV
X-Y
√
Z
√
A
X-Y
√
√
Calibration of “A” Scanners for Atomic-scale
Z
√
200nm deep
10µm
10µm