VECTOR ™- USMC VECTOR 21-V.0.3
38
Settings
Calibration procedure (hand held)
Point the VECTOR roughly
northwards.
Click the azimuth key four times
in rapid succession. ”FIEL Co”
appears briefly, followed by
”4 Pt Co” for the regular 4 point
calibration.
If you want to perform the
12 point calibration click
the distance key when ”4 Pt Co”
is displayed until the desired
option is displayed. The
selected procedure starts
automatically in a few moments.
Move the VECTOR according to
the displayed instructions (see
page 37).
After the last instruction,
analysis begins and the pointing
circle blinks for a few seconds.
You then see
• ”Good Co” followed by
accuracy value
• ”bAd Co” followed by accuracy
value
• ”ACC”: accuracy value in units
selected
4x
i