VECTECH 498A Operation Manual Download Page 4

VECTECH498A Operation Manual 

Page 2 

1. Summary 

VECTECH498A  Charged  plate  Monitor  (CPM)  measures  the  voltage  potential  on  its  integrated  plate-electrode 

according  the  Fieldmill-influence-principal  and  it  is  suitable  to  measure  the  static  de lay  time  of  air  ionizer 

equipment.  The  unit  meets  the  requirements  of  the  American  standard  ESD-STM  3.1-2000.  VECTECH498A  is 

menu-driven  and  it  is  easy  to  operation  by  four  switches.  All  parameters  will  be  shown  in  the  LCD-display  for  a 

better orientation of the operation. It is a special device for testing the capability of the electrostatic eliminator. 

离 子 采 集 板

I n s u l a t io n

C h a r g e d  p l a t e

 

 

 

 

 

 

 

 

 

 

 

 

 

 

2. Parts 

Parts 

Quantity 

CHARGE PLATE Monitor 498A 

Special power adapter 

Special toolbox 

Shoring pole 

Ground cable 

Operation manual 

Guarantee card   

1.

 

Sensor 

2.

 

Handle 

3.

 

Shoring pole 

4.

 

Ground jack 

5.

 

Power jack 

Summary of Contents for 498A

Page 1: ...e Plate Monitor Operation manual Thank you for purchasing our charged plate monitor Please carefully read this manual before operating the system Store this manual in a safe easily accessible place for future reference ...

Page 2: ...3 4 Power supply instruction 3 4 1 turn on the unit 3 4 2 turn off the unit 3 4 3 Charge 4 4 4 backlight 4 5 Test instruction 4 5 1 Test operation 4 5 2 Voltage Function 5 6 Set work mode and parameters 6 7 Zero by software 7 8 View the saved data 7 9 TypeSet setting 8 10 PassSet setting 8 11 ReSet 10 Appendix 11 A Calibration 11 ...

Page 3: ...charging and keep certain distance to avoid the capacitance affect 4 To avoid the high voltage damage the charged plate monitor no charging outer type power supply than the voltage is bigger than 1100v 5 It must use the special power supply adapter which is supplied by our company 6 No using the unit in the environment where the humidity is greater 7 No discharge the unit except the professional p...

Page 4: ... STM 3 1 2000 VECTECH498A is menu driven and it is easy to operation by four switches All parameters will be shown in the LCD display for a better orientation of the operation It is a special device for testing the capability of the electrostatic eliminator 离 子 采 集 板 In su la tio n C h a rg ed p la te 2 Parts Parts Quantity CHARGE PLATE Monitor 498A 1 Special power adapter 1 Special toolbox 1 Shor...

Page 5: ...g and viewing data and it can save one hundred data 4 An individual way of setting and set the work parameters meeting the demand such as key tone alert tone 5 There are charge batteries inside and it can use without outer power supply 6 With no touching sensor and the test resistance is stable and precise 7 It is portable and easy to use 4 Power supply instruction 4 1 turn on the unit Outer speci...

Page 6: ...power supply if not the backlight is off The backlight of the LCD is off when the unit is in charging state 5 Test instruction 5 1 Test operation 1 Take out the unit from the toolbox 2 Ground the unit and connect the ground of unit with the ground of tested device 3 Turn on the unit refer to the 4 1 4 Press key into the waiting state the LCD will show the testing mode AUTO POSI NEGA BAL Press VIEW...

Page 7: ...tate M Press BACK key to the waiting state at any test work mode POSI Mode Testing process is same with the some process of AUTO mode A B C D I J K L M NEGA Mode Testing process is same with the some process of AUTO mode E F G H I J K L M BAL Mode Testing process is same with the some process of AUTO mode I J K L M For Example testing process of the AUTO mode is as following PassFunction and SaveF...

Page 8: ...enters into the parameter setting and it can set START initial test voltage END final test voltage BAL ion balance time a Set START click or long press or key about five seconds to change the parameters The parameters change as following 500V 550V 600V 950V 1000V 500V b Set END press ENTER key to save the START value And then click or long press or key about five seconds to change the parameters T...

Page 9: ... types about the saved data A and B Press ENTER key to switch them Switch the type of saved data A to B by pressing ENTER key A the testing mode and its parameter setting B the testing data 4 Exit the process of viewing data by pressing BACK to the waiting interface Note It can save the testing data when only the SaveFunction is on It can save one hundred data and the sequence is sort ascending fr...

Page 10: ...d when pressing keys OFF Close the function and not sound when pressing the keys 5 Alert Tone ON Give sound when testing OFF Close the function For example the setting of PassFunction Select PassFunction from menu typeset by pressing or key and then press ENTER key into PassFunction setting Select ON from menu PassFunction by pressing or key and then press ENTER key to check it and to the next men...

Page 11: ...ter POSI setting If select NO return to the AUTO setting 2 POSI Refer to the setting of PT positive decay time and BAL U balance voltage of AUTO setting 3 NEGA Refer to the setting of NT positive decay time and BAL U balance voltage of AUTO setting 4 BAL Refer to the setting of BAL U balance voltage of AUTO setting 4 Press BACK key to the up menu For example the setting of the submenu AUTO of menu...

Page 12: ...pressing and keys and then check the password by pressing ENTER key means 1 means 0 the password is 01101001 3 If the inputting password is error the LCD shows PWD Error and returns to the menu 5 ReSet 4 If the inputting password is correct the LCD shows ReSet If select YES the monitor shows Master Reset and runs the process of reset The LCD shows ReSet OK after successful resetting and return to ...

Page 13: ...ying voltage is 1000v Remark it may affect the zero value after calibrating the voltage linearity so the charged plate must be ground If the display is not 0 after first calibrating recalibrating as the above methods 2 Zero by hardware Warn Use hardware zeroing only when the zero value warps too much and cannot be zero by the software Methods 1 Press ENTER key into the voltage test function 2 Pres...

Reviews: