VECTECH 498A Operation Manual Download Page 10

VECTECH498A Operation Manual 

Page 8 

9. TypeSet setting 

select menu by pressing 

  or 

key

and then pressing ENTER to check the setting.   

1

 

There are four menus down the menu “TypeSet”

:①

SaveFunction 

PassFunction 

Key Tone 

Alert Tone

 

 

2.    SaveFunction 

ON

  After ending test, ask whether saving the test data or not on the menu. 

OFF

Close the function and not save the test data. 

 

3.    PassFunction 

ON

  When test data do not pass, the unit gives some hints.                                                                                                                                                                                                                                                                 

OFF

Close the function and not hint when the data do not pass. 

 

Note:   

 

In the testing process, compare the test data with the setting pass  parameters (set them at the menu “

PassSet

”). If 

the test data is not bigger than the pass parameter, the test data is up to grade, or else not pass. 

 

If the “alert tone”  is on, the unit  gives the alarm sound when the test data  is not  pass, or else, only with  glint  of 

the data not passing. 

 

4.    Key Tone 

ON

  Give sound when pressing keys. 

OFF

Close the function and not sound when pressing the keys. 

 

5.    Alert Tone 

ON

  Give sound when testing. 

OFF

Close the function 

 

For example: the setting of PassFunction. 

Select  “PassFunction”  from  menu  “typeset”  by  pressing 

or

key  and  then  press  ENTER key  into  “PassFunction” 

setting. 

Select “ON” from menu “PassFunction” by pressing 

or

key and then press ENTER key to check it and to the next 

menu setting. 

press BACK key to exit this menu setting and it will return to the waiting state if continually pressing BACK key.   

 

 

 

 

 

10. PassSet setting   

1.  It  can  set  threshold  values  of  the  test  voltage  and  decay  time  under  the  submenus  of  the  menu  “PassSet”:

 

ATUO 

POSI 

NEGA

 

BAL.

 

2. Select “PassSet” from menu “MENU” by pressing 

or

key and then press ENTER key into “PassSet” setting. 

3. Select test mode as demanding and then press ENTER key into the parameter setting.  

1) AUTO

 

a.

 

Set 

“PT” positive decay time

, click or press 

or

key about five seconds to change the parameter. Parameter 

changes as following: 

TypeSet 

2.PassFunction

 

TypeSet 

3.Key Tone 

PassFunction 

1.ON

 

Press ENTER 

Press ENTER 

Summary of Contents for 498A

Page 1: ...e Plate Monitor Operation manual Thank you for purchasing our charged plate monitor Please carefully read this manual before operating the system Store this manual in a safe easily accessible place for future reference ...

Page 2: ...3 4 Power supply instruction 3 4 1 turn on the unit 3 4 2 turn off the unit 3 4 3 Charge 4 4 4 backlight 4 5 Test instruction 4 5 1 Test operation 4 5 2 Voltage Function 5 6 Set work mode and parameters 6 7 Zero by software 7 8 View the saved data 7 9 TypeSet setting 8 10 PassSet setting 8 11 ReSet 10 Appendix 11 A Calibration 11 ...

Page 3: ...charging and keep certain distance to avoid the capacitance affect 4 To avoid the high voltage damage the charged plate monitor no charging outer type power supply than the voltage is bigger than 1100v 5 It must use the special power supply adapter which is supplied by our company 6 No using the unit in the environment where the humidity is greater 7 No discharge the unit except the professional p...

Page 4: ... STM 3 1 2000 VECTECH498A is menu driven and it is easy to operation by four switches All parameters will be shown in the LCD display for a better orientation of the operation It is a special device for testing the capability of the electrostatic eliminator 离 子 采 集 板 In su la tio n C h a rg ed p la te 2 Parts Parts Quantity CHARGE PLATE Monitor 498A 1 Special power adapter 1 Special toolbox 1 Shor...

Page 5: ...g and viewing data and it can save one hundred data 4 An individual way of setting and set the work parameters meeting the demand such as key tone alert tone 5 There are charge batteries inside and it can use without outer power supply 6 With no touching sensor and the test resistance is stable and precise 7 It is portable and easy to use 4 Power supply instruction 4 1 turn on the unit Outer speci...

Page 6: ...power supply if not the backlight is off The backlight of the LCD is off when the unit is in charging state 5 Test instruction 5 1 Test operation 1 Take out the unit from the toolbox 2 Ground the unit and connect the ground of unit with the ground of tested device 3 Turn on the unit refer to the 4 1 4 Press key into the waiting state the LCD will show the testing mode AUTO POSI NEGA BAL Press VIEW...

Page 7: ...tate M Press BACK key to the waiting state at any test work mode POSI Mode Testing process is same with the some process of AUTO mode A B C D I J K L M NEGA Mode Testing process is same with the some process of AUTO mode E F G H I J K L M BAL Mode Testing process is same with the some process of AUTO mode I J K L M For Example testing process of the AUTO mode is as following PassFunction and SaveF...

Page 8: ...enters into the parameter setting and it can set START initial test voltage END final test voltage BAL ion balance time a Set START click or long press or key about five seconds to change the parameters The parameters change as following 500V 550V 600V 950V 1000V 500V b Set END press ENTER key to save the START value And then click or long press or key about five seconds to change the parameters T...

Page 9: ... types about the saved data A and B Press ENTER key to switch them Switch the type of saved data A to B by pressing ENTER key A the testing mode and its parameter setting B the testing data 4 Exit the process of viewing data by pressing BACK to the waiting interface Note It can save the testing data when only the SaveFunction is on It can save one hundred data and the sequence is sort ascending fr...

Page 10: ...d when pressing keys OFF Close the function and not sound when pressing the keys 5 Alert Tone ON Give sound when testing OFF Close the function For example the setting of PassFunction Select PassFunction from menu typeset by pressing or key and then press ENTER key into PassFunction setting Select ON from menu PassFunction by pressing or key and then press ENTER key to check it and to the next men...

Page 11: ...ter POSI setting If select NO return to the AUTO setting 2 POSI Refer to the setting of PT positive decay time and BAL U balance voltage of AUTO setting 3 NEGA Refer to the setting of NT positive decay time and BAL U balance voltage of AUTO setting 4 BAL Refer to the setting of BAL U balance voltage of AUTO setting 4 Press BACK key to the up menu For example the setting of the submenu AUTO of menu...

Page 12: ...pressing and keys and then check the password by pressing ENTER key means 1 means 0 the password is 01101001 3 If the inputting password is error the LCD shows PWD Error and returns to the menu 5 ReSet 4 If the inputting password is correct the LCD shows ReSet If select YES the monitor shows Master Reset and runs the process of reset The LCD shows ReSet OK after successful resetting and return to ...

Page 13: ...ying voltage is 1000v Remark it may affect the zero value after calibrating the voltage linearity so the charged plate must be ground If the display is not 0 after first calibrating recalibrating as the above methods 2 Zero by hardware Warn Use hardware zeroing only when the zero value warps too much and cannot be zero by the software Methods 1 Press ENTER key into the voltage test function 2 Pres...

Reviews: