22
Note:
Digital Video Monitor uses WIFI direct connection to connect to mopria-supported WIFI printers for
printing. The expected communication data involves medical and health data.
Digital Video Monitor can be connected to the local Internet through the RJ45 interface to obtain the
network synchronization time. The intended communication involves the interaction of network
synchronized time data.
7.
Cleaning and Disinfection
The product should be cleaned and disinfected before and after each use. It is recommended to clean and
disinfect the product according to the following instructions before and after use. Vathin Medical has
validated these instructions according to AAMI TIR 12 and 30. Any deviations from the instructions shall be
assessed properly to determine its effectiveness and potential adverse consequences to ensure that it
continues to achieve its intended purpose.
Warning
Disconnect Digital Video Monitor from any mains power supply, remove any accessories and make sure
the Digital Video Monitor is completely turned off before cleaning and disinfection.
You should take the following steps to clean and disinfect the product according to good medical practices:
7.1.
Cleaning
1.
Recommended detergent: enzymatic, mild pH of 7–9, and low foam (Enzol or equivalent).
2.
Prepare the cleaning solution by using the standard enzymatic detergent prepared according to the
manufacturer's recommendations.
3.
Dip the sterile gauze into the enzyme solution and ensure that the gauze is moist without dripping.
4.
Use the wet gauze to thoroughly clean the buttons and housing of the product. Well protect the product
from being wet to avoid damage to its internal electronic components.
4. Use a sterile soft brush dipped in enzyme solution to clean the buttons until all dirt is removed.
5. Wait 10 minutes (or the duration recommended by the detergent manufacturer) for the enzyme to be
activated.
6. Use the sterile gauze dipped in RO/DI water to clean the product to ensure that all traces of detergent
are removed.