5.1.4 Launching the exposure
To launch the X-ray emission, do the following steps.
1. Leave the X-Ray room and close the door. From outside of the x-ray room, you must
consistently monitor and pay attention to the patient during acquisition.
2. Launch X-ray using the exposure switch. Press and hold it until the end of acquisition.
Observe the LED lamp on top of the system, as it should turn orange, indicating X-ray
emission.
USB itself is not provided with the PaX-Primo i. Thus it is recommended
that the USB with following specification be mandatory to use.
Type : USB Flash Drive
Write Specification : USB 2.0
In cases of any problems, release the exposure switch button or press the
red emergency stop button on the column. (See the following figure)