background image

4. Product Specifications 

60 

EzRay Air Wall (Model: VEX-S300W) Installation Manual 

4.2 

Technical Specifications 

X-ray Generator

 

Item 

Description 

High Voltage 

Generator 

(Assembly) 

Model 

DG- S0102V1  

Rated output power 

Max. 0.2 kW  

Duty Cycle 

1:60 or more               

(Exposure time: Interval time)  

Cooling Protection 

Thermistor ≥ 65

 °C  

Inherent Filtration 

1.8 mm Al / 65 kV  

Total Filtration 

Min. 1.5 mm Al  

Type 

Inverter Type  

Tube Voltage 

55-65 kV  

Tube Current 

1.0-3.0 mA  

X-ray Tube 

Manufacturer 

VATECH Co., Ltd.  

Model 

V1-650304 (Stationary Anode type)  

Version 

2.0  

Focal spot size 

0.4 mm (IEC 60336)  

Anode heat contents 

Max. 2.7 kJ  

Maximum Anode Heat 

Dissipation 

200 W  

Target Material 

Tungsten  

Target Angle 

12.5°  

Inherent Filtration 

Min. 1.5 mm Al  

X-ray Coverage 

70 mm at SID 200 mm  

Tube Voltage 

Max. 65 kV  

Tube Current 

Max. 3.0 mA 

Summary of Contents for EzRay Air Wall VEX-S300W

Page 1: ......

Page 2: ......

Page 3: ......

Page 4: ......

Page 5: ...eview of this manual is recommended before installation to ensure proper installation of this equipment The EzRay Air Wall is in steady improvement The information contained in this manual may be subject to change without notice justification or notification of the persons concerned All brand names and logos used in this manual are copyrighted For further information not covered in this manual or ...

Page 6: ...equipment is used to its full potential Radiation symbols indicate a possible danger from exposure to radiation ESD susceptibility symbols indicate that an item is susceptible to damage from electrostatic discharges Warnings Regarding X ray Radiation Failure to install this equipment in an approved location may be dangerous to the patient and operator Stationary radiation shielding must be install...

Page 7: ...ly perform installation and related work Do not connect any items or equipment to this system which are not part of the system IEC60601 1 1 Any equipment not approved by VATECH must comply with the applicable standards IEC 60950 1 for IT equipment ex PC and IEC 60601 1 for medical electrical equipment All operators of this equipment are responsible for ensuring that the requirements outlined in IE...

Page 8: ...evices Do not install the equipment in an area that is exposed to strong electromagnetic fields Do not install this system in an area where there is the risk of an explosion The electrical installation of this system shall comply with all local code requirements for electro medical systems IEC 60364 7 710 2002 It is strongly recommended that a UPS be installed at the same time as the equipment Thi...

Page 9: ...s 7 3 Installing the Equipment 20 3 1 Marking Drill points on the wall 23 3 2 Installing the Power Box Assembly 24 3 3 Installing the Horizontal Arm 34 3 4 Installing the Scissor Arm 37 3 5 Installing the X Ray Generator Assembly 39 3 6 Connecting the Cables 43 4 Product Specifications 55 4 1 Mechanical Specifications 55 4 2 Technical Specifications 60 4 3 Electrical Specifications 61 4 4 Environm...

Page 10: ...Table of Contents vi EzRay Air Wall VEX S300W Installation Manual This page intentionally left blank ...

Page 11: ...a qualified technician s from one of our authorized agents The equipment has been used under a normal condition by the user s manual 1 2 Customer s Responsibility Site planning and preparation are the responsibility of the customer The following points should be considered fundamentally important to all customers of this product Install all required materials before delivery of the system Complete...

Page 12: ...1 Introduction 2 EzRay Air Wall Model VEX S300W Installation Manual This page intentionally left blank ...

Page 13: ...o install the equipment safely 2 1 Required Tools The following tools are necessary to install the equipment Item Figure Size Wrench Set 1 5 mm 10 mm Monkey Spanner 200 mm Ratchet with Extension Bar and Sockets Socket 13 mm for M8 17 mm for M10 Crosshead screwdriver with a magnetic tip L 200 mm Spirit Level Type 70 300 Anti static Gloves Knife Tape Ruler 5 m ...

Page 14: ...2 Before Installing the Equipment 4 EzRay Air Wall Model VEX S300W Installation Manual Item Figure Size Marker Pen thick tip Hammer Multi meter Hammer Drill Cutting Pliers ...

Page 15: ...ge If it appears damaged do not open the package and immediately contact the shipping company agent or VATECH Wear the anti static or general purpose gloves Refer to Parts and Accessories List on page 15 19 Anti static Gloves Part No 45 Qty 1 ea Package Size Item Size mm inch Weight kg lbs Length 1 137 mm 44 8 26 6 kg 58 6 lbs Horizontal Arm 900 mm Width 443 mm 17 4 Height 416 mm 16 4 h w l ...

Page 16: ...del VEX S300W Installation Manual Unpacking the Main Box 1 Move the main box to a convenient place as close as possible to the installation location 2 Remove each box in the following order A Top Cover B Upper Box C Middle Box D Lower Box A B C D A B C D ...

Page 17: ... 3 Checking the Components Upper Box B No Items Figure Qty Comments Confirmed 1 Wall Bracket Double Stud Double Stud Frame 1 YES NO Double Stud Cover 2 YES NO Drilling Template for Double Stud 1 Paper Size A2 YES NO Wall Bracket Single Stud Drilling Template for Single Stud 1 Paper Size A3 YES NO ...

Page 18: ...2 Wall Bracket Double Stud Fisher Anchor M10 x L40 6 YES NO Fisher Bolt M10 x L35 6 YES NO Wall Bracket Single Stud Blank 3 Wall Bracket Double Stud HEX HEAD WOOD SCREW M10 x L70 4 YES NO HEX HEAD WOOD SCREW M8 x L60 with Flat Washer M8 2 YES NO 2 YES NO Wall Bracket Single Stud Blank 4 Installation Manual 1 YES NO Anti static Gloves 1 YES NO 5 User Manual 1 YES NO ...

Page 19: ...l VEX S300W Installation Manual 9 ENGLISH Middle Box C No Items Figure Qty Comments Confirmed 6 Horizontal Arm Assembly 1 3 options 1 450 mm 2 600 mm 3 900 mm YES NO 7 Accessory Kit 1 Rectangular Cover 2x3 option 1 YES NO 2 Rectangular Cover 4x3 1 YES NO ...

Page 20: ...Wrench Bolt M5 x L25 2 YES NO 4 X ray Generator Stopper 1 YES NO 5 X ray Generator Arm Bushing 1 YES NO 6 Truss Screw M4 x L6 7 YES NO Star Washer M4 3 YES NO 7 Generator Cover Bolt PH M3 x L6 3 YES NO 8 Fixing Truss Screw for X ray Generator Stopper M3 x L4 1 YES NO 9 Silicone Cap White 2 YES NO 10 Silicone Cap White 2 YES NO 11 Power Box Sticker 1 YES NO 12 X ray Generator Arm Screw Cover 3 1 YE...

Page 21: ...m Screw Cover 1 1 YES NO 15 ANTI ROTATION BLOCK 2 1 ACETAL Black YES NO ANTI ROTATION BLOCK STIFFENER 1 Spc YES NO 16 Fisher Anchor for Single Stud M10 x L40 3 For Single Stud YES NO Fisher Bolt for Single Stud M10 x L35 3 For Single Stud YES NO 17 Cable Tie 5 YES NO 8 X ray Generator Arm Cover 1 YES NO Power Box Cover Deco 1 YES NO Horizontal Arm Cable Cover 1 YES NO Horizontal Arm Frame Cover 2 ...

Page 22: ...igure Qty Comments Confirmed 9 Wrench Bolt for Double Stud M8 x L20 6 For Double Stud YES NO Double Stud Cover Screw FH M4 x L8 8 For Double Stud YES NO Double Stud Sticker 2 For Double Stud YES NO 10 AC Power Cable Option 1 Length 2 m YES NO 11 Power Box Assembly including Power Box Cover 1 YES NO ...

Page 23: ...x D No Items Figure Qty Comments Confirmed 12 Scissor Arm Assembly including cables 1 YES NO 13 X ray Generator Assembly including cables and Round Cover 1 YES NO 14 Remote Exposure Switch Assembly including Holder and Double sided Sticker 1 YES NO Remote Exposure Switch Holder Screws FH M3 x L16 2 YES NO ...

Page 24: ...tch Cable Doorbell type Standard or Option 1 Standard for the USA Option for the other regions YES NO 17 Remote Exposure Switch Doorbell type Standard or Option 1 Standard for the USA Option for the other regions YES NO C S Component Box Optional No Items Figure Qty Comments Confirmed 1 Rotating Rectangular Cover 2x3 3x2 Option 1 YES NO 2 Rotating Rectangular Cover 4x3 3x4 Option 1 YES NO ...

Page 25: ... ABS White 2 YES NO 2 Horizontal Arm Cable Cover Material ABS White 1 YES NO 3 Power Box Cover Deco Material ABS White 1 YES NO 4 Silicone Cap white Material Silicone 2 YES NO 5 Truss Screw Size M4 x L6 7 YES NO 6 Star Washer Size M4 3 YES NO 7 X ray Generator Arm Bushing Material AL 1 YES NO 8 X ray Generator Stopper 1 YES NO 9 X ray Generator Arm Cover 1 YES NO 10 X ray Generator Arm Screw Cover...

Page 26: ...ator Cover Bolt Size PH M3 x L6 3 YES NO 14 HEX HEAD WOOD SCREW Single Stud Size M10 x L70 3 YES NO 15 Fisher Anchor Single Stud Size M10 x L40 3 YES NO 16 Fisher Bolt Single Stud Size M10 x L35 3 YES NO 17 Drilling Template Single Stud Paper Size A3 Single Stud 1 YES NO 18 Fixing Truss Screw for X ray Generator Stopper Size M3 x L4 1 YES NO 19 ANTI ROTATION BLOCK 2 Material ACETAL Black 1 YES NO ...

Page 27: ...Bolt Size M5 x L25 2 YES NO 22 Remote Exposure Switch Assembly including Holder and Double sided Sticker 1 YES NO 23 Remote Exposure Switch Holder Screws Size FH M3 x L16 2 YES NO 24 AC Power Cable Option Length 2 m 1 YES NO 25 Drilling Template Double Stud Paper Size A2 Double Stud 1 YES NO 26 Cable Tie 3 YES NO 27 Rectangular Cover 2x3 Material ABS FOV 2x3 3x2 1 YES NO ...

Page 28: ...sure Switch Doorbell type Standard or Option Standard for the USA Option for the other regions 1 YES NO 31 Remote Exposure Switch Cable Doorbell type Standard or Option Standard for the USA Option for the other regions 1 YES NO 32 Door Interlock Cable Option 1 YES NO 33 Power Box Sticker 1 YES NO 34 Double Stud Frame 1 YES NO 35 Double Stud Cover Material ABS White 2 YES NO 36 Double Stud Sticker ...

Page 29: ...X HEAD WOOD SCREW Double Stud Size M8 x L60 2 YES NO 40 Flat Washer Double Stud Size M8 2 YES NO 41 Fisher Anchor Double Stud Size M10 x L40 6 YES NO 42 Fisher Bolt Double Stud Size M10 x L35 6 YES NO 43 Wrench Bolt Double Stud Size M8 x L20 6 YES NO 44 Double Stud Cover Screw Size FH M4 x L8 8 YES NO 45 Anti static Gloves 1 YES NO 46 Rotating Rectangular Cover 2x3 Material ABS FOV 2x3 3x2 1 YES N...

Page 30: ...2 Before Installing the Equipment 20 EzRay Air Wall Model VEX S300W Installation Manual This page intentionally left blank ...

Page 31: ...nting supports must be used to fulfill this requirement You are advised to plan and study the installation environment carefully in advance before proceeding since the installation involves drilling the wall Accurate measurement is of critical importance for a successful installation The horizontal arm comes with one of three different lengths depending on your purchase L 450 600 or 900 mm Perform...

Page 32: ...le Stud Wood wall Wood screw M10 x L70 X 3 Part No 14 Concrete wall Fisher Anchor M10 x L40 X 3 No 15 Fisher Bolt M10 x L35 X 3 No 16 Wall Bracket Installation Double Stud Wood wall Wood screw M10 x L70 X 4 Part No 38 Wood screw M8 x L60 X 2 Part No 39 Flat Washer M8 X 2 Part No 40 Concrete wall Fisher Anchor M10 x L40 X 6 Part No 41 Fisher Bolt M10 x L35 X 6 Part No 42 Part No 38 Part No 39 40 ...

Page 33: ...mplate Single Stud Part No 17 or Double Stud Part No 25 temporally Place it on the wall and ensure leveling is obtained the bubble should be located in the middle of the spirit level 3 If it is tape the drilling template temporally at three or four corners 4 Mark the holes on the wall according to the mounting type and then remove the drilling template Single Stud Double Stud ...

Page 34: ...mbly 3 2 1 Mounting the Power Box Assembly For wood walls Cabinet Installation and Wall Mount Installation Type Single Stud Ratchet with Extension Bar and Socket Socket Size 17 mm Crosshead screwdriver w magnetic tip L 200 mm HEX HEAD WOOD SCREW Part No 14 Size M10 x L70 Qty 3 ea Power Box Assembly Cables included ...

Page 35: ...Installation Manual 25 ENGLISH 1 Remove the Power Box Cover of the Power Box Assembly 2 Attach the Power Box Base to the wall with three wood screws Ratchet with Extension Bar and Socket Socket Size 17 mm HEX HEAD WOOD SCREW Part No 14 Size M10 x L70 Qty 3 ea ...

Page 36: ...ep 1 on page 25 2 Attach the Double Stud Frame Part No 33 to the wall with six wood screws Ratchet with Extension Bar and Socket Socket Size 17 mm for M10 13 mm for M8 HEX HEAD WOOD SCREW Part No 38 Size M10 x L70 Qty 4 ea 1 Wood Screw Part No 39 Size M8 x L60 Qty 2 ea 2 Flat Washer Part No 40 Size M8 Qty 2 ea Part No 38 Part No 39 40 ...

Page 37: ...el VEX S300W Installation Manual 27 ENGLISH 3 Assemble the Double Stud Covers Part No 35 with eight flat head screws Double Stud Cover Screw Part No 44 Size FH M4 x L8 Qty 8 ea 4 Attach two Double Stud Stickers Part No 36 on each Double Stud Cover ...

Page 38: ...g the Equipment 28 EzRay Air Wall Model VEX S300W Installation Manual 5 Attach the Power Box Base to the Double Stud Frame with six wrench bolts Allen Wrench Size 6 mm Wrench Bolt Part No 43 Size M8 x L20 Qty 6 ea ...

Page 39: ...mbly For Concrete walls Cabinet Installation and Wall Bracket Installation Type Single Stud 1 Drill the marked points using the hammer drill 2 Remove the debris and clean the holes using the dust pump 3 Put the Fisher Anchor Part No 14 into the holes using the hammer Hammer Fisher Anchor Part No 15 Size M10 x L40 Qty 3 ea ...

Page 40: ...tion Manual 4 Remove the Power Box Cover of the Power Box Assembly 5 Attach the Power Box Base to the anchors with three fisher bolts DO NOT tighten the bolts entirely until after leveling is complete Allen Wrench Size 8 mm Fisher Bolt Part No 16 Size M10 x L35 Qty 3 ea ...

Page 41: ...29 2 Put the Fisher Anchor into the holes using the hammer Hammer Fisher Anchor Part No 41 Size M10 x L40 Qty 6 ea 3 Attach the Double Stud Frame to the anchors with six fisher bolts DO NOT tighten the bolts entirely until after leveling is complete Allen Wrench Size 8 mm Fisher Bolt Part No 42 Size M10 x L35 Qty 6 ea 4 Perform Step 3 through 5 on page 27 and 28 ...

Page 42: ... four wrench bolts before placing the spirit level on the equipment Allen Wrench Size 3 mm 2 Place the spirit level on the position as shown in the figure 3 To adjust the horizontality level loosen one wrench bolt by using the wrench 4 Adjust the headless wrench bolt until the bubble on the spirit level is located in the middle Allen Wrench Size 3 mm ...

Page 43: ...e bolt tightened they are required to be loosened prior to leveling it 3 2 4 Leveling the Power Box Assembly Left Right 1 Place the spirit level on the position as shown in the figure 2 After placing the spirit level on its position adjust the two bolts until the bubble on the spirit level is located in the middle Allen Wrench Size 3 mm ...

Page 44: ...he arm s length 1 After pulling out the cables through the opening insert the horizontal arm into the hole on the Power Box Assembly 2 Fix the block assembly using the anti rotation block2 anti rotation block stiffness and wrench bolts Allen Wrench Size 4 mm Horizontal Arm Cable included ANTI ROTATION BLOCK2 Part No 19 Qty 1 ea ANTI ROTATION BLOCK STIFFENER Part No 20 Qty 1 ea Wrench Bolt Part No ...

Page 45: ...H Make sure that the Horizontal Arm moves smoothly If not slightly loosen the wrench bolts 3 3 1 Leveling the Horizontal Arm 1 Place the spirit level on the position as shown in the following figures Front Rear Left Right 2 Check if the bubble on the spirit level is located in the middle ...

Page 46: ...n the middle follow the same procedures in 3 2 3 Leveling the Power Box Assembly Front Rear and 3 2 4 Leveling the Power Box Assembly Left RightError Reference source not found After leveling the Horizontal Arm is complete make sure to fasten the four wrench bolts as shown in the following figures ...

Page 47: ...move the safety strap and shrink wrap for scissor arm until the installation is completed When you remove the strap and shrink wrap the scissor arm will be opened by its spring tension It can cause injury to personnel or damage to equipment 1 Insert the cable and the scissor arm into the opening of the horizontal arm Scissor Arm Assembly Cable included Safety Shrink Strap Wrap ...

Page 48: ...ual 2 Tighten the wrench bolt in the Horizontal Arm Frame End and cover the Frame End with the Horizontal Arm Frame Cover Part No 1 Horizontal Arm Frame Cover Part No 1 Qty 2 ea Location of the hole to tighten the wrench bolt Tightening the wrench bolt and covering the Frame End ...

Page 49: ...embly Handle the X Ray tube head with care when installing the X Ray tube head 1 Place X ray Generator Arm Bushing Part No 7 on the top of the hole in the X Ray Generator Arm 2 Attach the X ray Generator Assembly to the scissor arm while making sure that the cables from the Scissor Arm go through the X Ray Generator Arm s hole ...

Page 50: ...nnecting cables please see 3 6 4 Cabling between the Scissor Arm and the X ray Generator on page 48 4 Attach the X ray Generator Stopper Part No 8 to the rear top of the X ray Generator Arm 5 Tighten the screw above the attached stopper Truss Screw for X ray Generator Stopper Part No 18 Size M3 x L4 Qty 1 ea ...

Page 51: ...S300W Installation Manual 41 ENGLISH 6 Cover the rear side of the X ray Generator Arm with the X ray Generator Arm Cover Part No 9 7 Tighten three bolts on the rear cover of the X ray Generator Arm Generator Cover Bolt Part No 13 Size PH M3 x L6 Qty 3 ea ...

Page 52: ...ment 42 EzRay Air Wall Model VEX S300W Installation Manual 8 Cover three holes with caps GENERATOR ARM COVER CAP1 Part No 10 Qty 1 ea GENERATOR ARM COVER CAP2 Part No 11 Qty 1 ea GENERATOR ARM COVER CAP3 Part No 12 Qty 1 ea ...

Page 53: ...t No 45 2 DO NOT wear a thick jacket Always ensure that the cables are firmly connected A bad connection could cause deterioration of image quality 3 6 1 Connecting the Power Cord 1 Connect the N Neutral and L Live Connectors of the permanent power cord to the terminal blocks and Protective Earth PE Cable to the Protective Earth Conductor L Live Brown N Neutral Light Blue Protective Earth PE Cable...

Page 54: ...res are the same regardless of the Horizontal Arm s length 1 Connect the cable H001624 with the Connector 1 2 on the Power Box Board and the Protective Earth PE Cable on the PE Conductor When you connect the Protective Earth PE Cables to the PE conductor by using the Truss Screw Part No 5 with Star Washer Part No 6 please refer to the illustration below CN Connector 2 CN Connector 1 Protective Ear...

Page 55: ...ss screws on the Power Box Deco Truss Screw Part No 5 Size M4 x L6 Qty 2 ea 3 Tighten two truss screws on the bottom of the Power Box Cover Truss Screw Part No 5 Size M4 x L6 Qty 2 ea 4 If the device has been installed with the permanent power cord attach the Power Box Sticker to the Connector 4 Power Box Sticker Part No 32 Qty 1 ea Power Box Deco Connector 4 ...

Page 56: ...Model VEX S300W Installation Manual 3 6 3 Cabling between the Scissor arm and horizontal arm During cable installation avoid any cable kinks and maintain the proper bending Ensure that cables are not caught on the cover Horizontal Arm Scissor Arm ...

Page 57: ... and scissor arm 1 White connector 3 pin White connector 3 pin 2 Black connector 8 pin Black connector 8 pin 3 Protective Earth PE Cable from the Scissor Arm When you connect the Protective Earth PE Cables to the PE conductor by using the Truss Screw Part No 5 with Star Washer Part No 6 please refer to the illustration below 1 2 3 ...

Page 58: ...erial ABS White Part No 2 Qty 1 ea 3 6 4 Cabling between the Scissor Arm and the X ray Generator 1 White connector 3 pin White connector 3 pin 2 Black connector 8 pin Black connector 8 pin 3 Protective Earth PE Cable from the Scissor Arm When you connect the Protective Earth PE Cables to the PE conductor by using the Truss Screw Part No 5 with Star Washer Part No 6 please refer to the illustration...

Page 59: ...3 Installing the Equipment EzRay Air Wall Model VEX S300W Installation Manual 49 ENGLISH 3 6 5 Tie the cables 1 Tie the cables together with a cable tie Cable Tie Part No 26 Qty 2 ea ...

Page 60: ...3 Installing the Equipment 50 EzRay Air Wall Model VEX S300W Installation Manual 3 6 6 View after Completion ...

Page 61: ...ard and optional Remote Exposure Switches and the Door Interlock Cable are connected to the Connector 1 2 and 3 Please check all seven possible options in the table below Before connecting the cables ensure that the DIP switch setting is set according to the instructions in the following table Refer to the location of the DIP switch in the following figure Option No Description Connector 1 Connect...

Page 62: ...e Exposure Switch Standard 2 Remote Exposure Switch Option Option 3 Press 1 and 2 at the same time N A AC Power Cable Option 1 Remote Exposure Switch Standard 2 Remote Exposure Switch Doorbell type Option Option 4 Press 1 N A AC Power Cable Option 1 Remote Exposure Switch Standard 2 Door Interlock Cable Option Option 5 Press 1 or 2 N A AC Power Cable Option 1 Remote Exposure Switch Standard 2 Remo...

Page 63: ...nnector 4 Option 6 Press 1 or 2 N A AC Power Cable Option 1 Remote Exposure Switch Standard 2 Remote Exposure Switch Doorbell type Option Option 7 Press 2 and 3 at the same time 1 Not used in this option AC Power Cable Option 1 Remote Exposure Switch Standard 3 Remote Exposure Switch Doorbell type Option 2 Remote Exposure Switch Doorbell type Option ...

Page 64: ...3 Installing the Equipment 54 EzRay Air Wall Model VEX S300W Installation Manual ...

Page 65: ...4 Product Specifications EzRay Air Wall Model VEX S300W Installation Manual 55 ENGLISH 4 Product Specifications 4 1 Mechanical Specifications Dimensions ...

Page 66: ...4 Product Specifications 56 EzRay Air Wall Model VEX S300W Installation Manual ...

Page 67: ...4 Product Specifications EzRay Air Wall Model VEX S300W Installation Manual 57 ENGLISH ...

Page 68: ...4 Product Specifications 58 EzRay Air Wall Model VEX S300W Installation Manual With Horizontal Arm 900 mm With Horizontal Arm 600 mm With Horizontal Arm 450 mm ...

Page 69: ...h Scissor Arm 900 mm Folded 1077 0 L x892 6 H x58 4 W 11 45 Unfolded 2338 7 L x327 0 H x58 4 W 600 mm Folded 777 0 L x892 6 H x58 4 W 10 6 Unfolded 2038 7 L x327 0 H x58 4 W 450 mm Folded 627 0 L x892 6 H x58 4 W 10 15 Unfolded 1888 7 L x327 0 H x58 4 W Horizontal Arm Assembly 900 mm 976 0 H x 151 2 V x 58 3 W 3 65 600 mm 676 0 H x 151 2 V x 58 3 W 2 8 450 mm 526 0 H x 151 2 V x 58 3 W 2 35 Scisso...

Page 70: ...hermistor 65 C Inherent Filtration 1 8 mm Al 65 kV Total Filtration Min 1 5 mm Al Type Inverter Type Tube Voltage 55 65 kV Tube Current 1 0 3 0 mA X ray Tube Manufacturer VATECH Co Ltd Model V1 650304 Stationary Anode type Version 2 0 Focal spot size 0 4 mm IEC 60336 Anode heat contents Max 2 7 kJ Maximum Anode Heat Dissipation 200 W Target Material Tungsten Target Angle 12 5 Inherent Filtration M...

Page 71: ...gle phase Power Rating 4 2 A Tube Voltage 65 kV fixed 3 Tube Current Option A 2 5 mA Option B 3 0 mA Accuracy Error kVp 10 mA 20 s 5 or 20 ms The input line voltage depends on the local electrical distribution system Allowable input voltage fluctuation requirement 10 Mode of operation Continuous operation with cyclic loading 60 times the exposure time wait before next exposure can begin ...

Page 72: ...ation Manual 4 4 Environmental Specifications Item Description During operating Temperature 10 35 Relative humidity 30 75 Atmospheric pressure 860 1060 hPa Transport and storage Temperature 10 60 Relative humidity 10 75 non condensing Atmospheric pressure 860 1060 hPa ...

Page 73: ...Appendix EzRay Air Wall Model VEX S300W Installation Manual 63 ENGLISH Appendix Drilling Template Single Stud ...

Page 74: ...Appendix 64 EzRay Air Wall Model VEX S300W Installation Manual Double Stud ...

Page 75: ...how to connect the third party exposure switch with the equipment fromVATECH How to 1 Cut the exposure switch cable provided with the equipment 2 According to the following schematic diagram rewire the cables 3 Double check the wiring before use Note tape the end of each unused wire to prevent the wires from causing an inadvertent short circuit ...

Page 76: ...mation about the equipment purchaser Customer Name of Clinic or Hospital Address Phone E Mail Information about the equipment seller Dealer Name of dealer Phone E Mail Website 2 Installation information Installation information Names of installers Scheduled date of installation Date of installation Model Serial No ...

Page 77: ... to manual Did installer make sure harness and equipment are well connected and not damaged Did the equipment be well balanced 4 After installation After installation Checklist Yes No Is it OK after checking the equipment visually Does each arm move smoothly Do the arms horizontal and scissor drop down by itself Does the LED on the front of the equipment display usually Does the equipment supply t...

Page 78: ...t information contact your VATECH representative Manufactured by VATECH Co Ltd Tel 82 1588 9510 Email gcs vatech co kr Website www vatech co kr Head Quarters Address 13 Samsung 1 ro 2 gil Hwaseong si Gyeonggi do 18449 Korea Factory Address 13 Samsung 1 ro 2 gil Hwaseong si Gyeonggi do 18449 Korea The CE symbol grants this product compliance to the European Directive for Medical Devices 93 42 EEC a...

Page 79: ......

Page 80: ...Copyright VATECH Networks All Rights Reserved ...

Reviews: