2.3.
CT Sensor Calibration
To ensure CT image quality, perform CT sensor calibration according to this section.
To assure image quality, perform sensor calibration in various temperature
and humidty settings.
2.3.1.
CT Sensor Calibration Standard
See the tables below for dark and bright calibration standards.
Dark Calibration Median Value Standard
Min
Max
Median value
700
2300
Bright Calibration Standard Value
Acquisition conditions
Half Scan
4-point
2-point
Filter
AL filter
Al
X
Cu filter(Cu 1 mm)
O
X
※
Using filter by each calibration point
Ensure that the median value for each bright cal point meets the standards in the table below:
Bright Standard Values: 2-point without Cu 1 mm filter
Bright
Min
Max
Median value
Cal Point 1
4500
5500
Cal Point 2
9500
10500
Bright Standard Values: 4-point with Cu 1 mm filter
Bright
Min
Max
Median value
Cal Point 1
80
120
Cal Point 2
150
700
Cal Point 3
800
1800
Cal Point 4
2000
2500
After bright calibration,
Cal point 1
and
4
must
satisfy the standards listed
above. The date from other Cal points can be different from the standards
because of the sensor you use for the calibration.
Summary of Contents for A9 PHT-30CSS
Page 1: ...vatech A9 Technical manual Model PHT 30CSS Version 1 01 English...
Page 12: ...Page 11 of 244 1 2 1 Vertical Frame Part...
Page 22: ...Page 21 of 244 1 2 3 5 Generator DG 07E22T2...
Page 26: ...Page 25 of 244...
Page 27: ...Page 26 of 244...
Page 30: ...Page 29 of 244 1 3 3 Sensor Part Connection Diagram...
Page 31: ...Page 30 of 244 1 3 4 CEPH SENSOR Part Connection Diagram...
Page 32: ...Page 31 of 244 1 3 5 Connection Diagram Total assembly...
Page 43: ...Page 42 of 244 16 Loosen ten bolts as shown in the image below...
Page 112: ...Page 111 of 244 5 Click the V Vol button Click the Ready button...
Page 180: ...Page 179 of 244 10 Select 50 in the M box and click the center of the flat washer...
Page 209: ...Page 208 of 244 8 5 Appendix 8 5 1 Main MCU Board Parameters...
Page 244: ...vatech A9 www vatech com...