2.8.
CT Number Calibration
To optimize CBCT image quality, the operator should perform the CBCT CT Number Calibration
according to the following procedures.
Make sure that the operator should use
TYPE A
as CT PHANTOM JIG
during the CT Number Calibration.
(
SAP Code
- Type A: M0103175 / Type B: M0104116)
Make sure that the operator should use
TYPE A
as CT NUMBER CHECK
PHANTOM during the CT Number Calibration.
(
SAP Code
-Type A: M0103175 / Type B: M0104116)
<Type A(>
<Type B>
<Type A>
<Type B>